FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 5437875 · Received February 16, 2016

Report

Report Number
3004123209-2016-00221
Event Type
Malfunction
Date Received
February 16, 2016
Date of Event
February 10, 2016
Report Date
April 13, 2016
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE USER FIRST ATTEMPTED TO INSTALL THE PAD-PAK ON THE 19TH JUNE 2013, THE PAD-PAK WAS SUCCESSFULLY INSTALLED A MINUTE LATER AND PASSED A SELF-TEST. THE DEVICE PERFORMED SELF-TESTS, ALONGSIDE RANDOM MANUAL POWER ONS, UP TO AND INCLUDING THE LAST LOG ENTRY ON THE 28TH FEBRUARY 2016. A TEST PAD-PAK WAS INSERTED INTO THE DEVICE AND THE DEVICE PASSED THE SUBSEQUENT SELF-TEST. NO STATUS LED WAS VISIBLE THROUGHOUT THE POWER CYCLE. THE DEVICE POWERED OFF WITHOUT ANY WARNINGS AND THE GREEN STATUS LED REMAINED EXTINGUISHED. THIS WOULD CONFIRM THE REPORTED FAULT. AN ACCEPTABLE SHOCK WAS DELIVERED DURING THE TESTING. INVESTIGATION FOUND THE REPORTED FAULT CAN BE ATTRIBUTED TO FAILURE OF THE SILVER PASTE CONNECTION ON GREEN STATUS LED. THE GREEN STATUS LED WAS TESTED DURING FINAL TEST, IT IS CONCLUDED THAT THE FAILURE WAS CAUSED AFTER THE DEVICE PASSED 'OUT QAT' ON THE 30TH APRIL 2013. THE FAULT COULD NOT BE REPLICATED WITH A NEW MEMBRANE FITTED.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. NO STATUS INDICATOR ON DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94853 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM N/A

Patients

Seq Age Sex Outcome Treatment
1