FDA Adverse Event Malfunction Summary report: N

TECNIS CL

MDR report key: 5437514 · Received February 15, 2016

Report

Report Number
2648035-2016-00243
Event Type
Malfunction
Date Received
February 15, 2016
Date of Event
January 18, 2016
Report Date
April 2, 2016
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
HQL
PMA / PMN Number
P880081
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

THE INTRAOCULAR LENS WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. VISUAL INSPECTION AT 10X MICROSCOPE MAGNIFICATION WAS PERFORMED AND ONLY A HALF OF LENS WAS RECEIVED. ONE LENS HAPTIC WAS OBSERVED DISTORTED. THE CONDITION OBSERVED IN THE LENS AND THE WAY IN WHICH THE CUT IS FORMED IS CONSISTENT WITH A LENS THAT HAS BEEN CUT DUE TO "IOL REMOVAL" DURING SAME SURGICAL PROCEDURE. THE COMPLAINT ISSUE AS IOL TORN (IOL OPTIC APPEARS CUT, RIPPED/TORN) WAS VERIFIED IN THE RETURNED UNIT. MANUFACTURING RECORDS WERE REVIEWED AND THE LENS WAS MANUFACTURED ACCORDING TO SPECIFICATION. NO ISSUES WERE REPORTED DURING THE MANUFACTURING PROCESS OF THIS PRODUCTION ORDER. A SEARCH ON COMPLAINTS REVEALED NO OTHER COMPLAINTS WERE RECEIVED FOR THIS ORDER NUMBER TO DATE. THE DIRECTIONS FOR USE (DFU) WAS REVIEWED. THE DFU ADEQUATELY PROVIDES INSTRUCTIONS AND PRECAUTIONS FOR THE PROPER USE AND HANDLING OF THE PRODUCT. BASED ON THE MANUFACTURING RECORDS REVIEW, ANALYSIS OF THE RETURN, HISTORICAL COMPLAINT REVIEW AND LABELING REVIEW, THERE IS NO INDICATION OF A PRODUCT MALFUNCTION OR PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CRACK WAS NOTED IN THE INTRAOCULAR LENS (IOL) AND IT WAS REMOVED FROM THE EYE DURING THE SAME SURGERY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93794 TECNIS CL MONOFOCAL IOLS HQL ABBOTT MEDICAL OPTICS Z9002

Patients

Seq Age Sex Outcome Treatment
1