COULTER ACT DIFF 2 ANALYZER
Report
- Report Number
- 1061932-2016-00157
- Event Type
- Malfunction
- Date Received
- February 10, 2016
- Date of Event
- January 19, 2016
- Report Date
- January 19, 2016
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K990352
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CORRECTION: THE DEVICE DATE OF MANUFACTURE WAS CHANGED FROM 08/01/2004 02/01/2005.
THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT ON 01/22/2016. AFTER TROUBLESHOOTING ANOTHER ISSUE THE FSE FOUND THAT CONTROLS FOR WBC WERE FAILING. THE FSE FOUND THAT THE WBC BATH WAS FAULTY. THE FSE REPLACED THE WBC BATH, WHICH REPAIRED THE FAILING WBC RESULTS ON CONTROLS. THE REPAIRS WERE VERIFIED BY THE FSE. THE BECKMAN COULTER INTERNAL IDENTIFIER FOR THIS EVENT IS (B)(6).
THE FIELD SERVICE ENGINEER (FSE) REPORTED THE COULTER ACT DIFF 2 ANALYZER WAS RECOVERING ERRONEOUS WHITE BLOOD CELL (WBC) RESULTS ON CONTROLS. ERRONEOUS PATIENT RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86293 | COULTER ACT DIFF 2 ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |