LIBERATOR 45
Report
- Report Number
- 3004972304-2016-00005
- Event Type
- Malfunction
- Date Received
- February 3, 2016
- Report Date
- May 17, 2016
- Manufacturer
- CAIRE INC.
- Product Code
- BYJ
- PMA / PMN Number
- K800742
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ONLY THE QDV WAS RETURNED TO THE COMPANY FOR NON-DESTRUCTIVE TESTING. THE QDV WAS BROKEN INTO TWO PIECES. THE BROKEN SECTION OF THE QDV WAS MEASURED AND THE ID AND OD WERE FOUND TO BE WITHIN SPECIFICATIONS. THE BROKEN QDV IS CONSISTENT WITH AN EXCESSIVE CANTILEVER LOAD LAVING BEEN APPLIED TO THE TOP SECTION OF THE PART. THE MOST LIKELY CAUSE OF THE FAILURE WAS AN EXCESSIVE BENDING FORCE APPLIED BY THE "FILLING STICK", WHICH IS ASSUMED TO BE THE MATING PORTION OF THE QDV WHICH IS CONNECTED TO THE LOX TRANSFER HOSE.
WE ARE ATTEMPTING TO HAVE THE UNIT RETURNED FOR EVALUATION. AFTER TESTING HAS BEEN COMPLETED A FOLLOW-UP REPORT WILL BE SUBMITTED.
THE COMPANY WAS NOTIFIED ON (B)(6) 2016 OF AN INCIDENT INVOLVING A HOME CARE PROVIDER TECHNICIAN. THE DATE OF THE INCIDENT IS UNKNOWN. THE INCIDENT WAS DESCRIBED AS FOLLOWS: THE "TECHNICIAN, IN HIS CAR, WAS FILLING HIS 45L UNIT, LINKED TO HIS 20000L CISTERN BEFORE HE LEFT ON TOUR. WHEN IT WAS FULL HE REMOVED THE FILLING STICK. AT THIS POINT A LARGE VIOLENT JET OF LIQUID OXYGEN CAME FROM THE TANK CONNECTOR. NARROWLY AVOIDING THE PROJECTION OF LIQUID OXYGEN, THE TECHNICIAN HASTILY MOVED TO THE OUTSIDE OF THE CARE, WHEN THE UNIT WAS EMPTY HE NOTICED THAT A PART OF THE CONNECTOR TO THE DEWAR WAS STUCK TO THE STICK, THE LATTER HAVING COLLAPSED." THERE WAS NO INJURY TO THE TECHNICIAN.
THE COMPANY WAS NOTIFIED ON JANUARY 8, 2016 OF AN INCIDENT INVOLVING A HOME CARE PROVIDER TECHNICIAN. THE DATE OF THE INCIDENT IS UNKNOWN. THE INCIDENT WAS DESCRIBED AS FOLLOWS: THE "TECHNICIAN, IN HIS CAR, WAS FILLING HIS 45L UNIT, LINKED TO HIS 20000L CISTERN BEFORE HE LEFT ON TOUR. WHEN IT WAS FULL HE REMOVED THE FILLING STICK. AT THIS POINT A LARGE VIOLENT JET OF LIQUID OXYGEN CAME FROM THE TANK CONNECTOR. NARROWLY AVOIDING THE PROJECTION OF LIQUID OXYGENT, THE TECHNICIAN HASTILY MOVED TO THE OUTSIDE OF THE CARE, WHEN THE UNIT WAS EMPTY HE NOTICED THAT A PART OF THE CONNECTOR TO THE DEWAR WAS STUCK TO THE STICK, THE LATTER HAVING COLLAPSED." THERE WAS NO INJURY TO THE TECHNICIAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66875 | LIBERATOR 45 | UNIT, LIQUID OXYGEN, STATIONARY | BYJ | CAIRE INC. | LIBERATOR 45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |