FDA Adverse Event Injury Summary report: N

CAPSURE Z NOVUS

MDR report key: 5420897 · Received February 9, 2016

Report

Report Number
2649622-2016-00444
Event Type
Injury
Date Received
February 9, 2016
Date of Event
December 11, 2015
Report Date
December 11, 2015
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY:THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. THE ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED LOWER RANGE. RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE TREND NOT AVAILABLE, HOWEVER AT INTERROGATION RV PACING IMPEDANCE MEASURED 192 OHMS. A 5061886 SHORT CIRCUIT COUNTS SINCE (B)(6) 2015. THE ANALYSIS OF THE DEVICE MEMORY INDICATED PACING CAPTURE THRESHOLD IN THE RV (RIGHT VENTRICLE) WAS ELEVATED. ON (B)(6) 2014, RV CAPTURE THRESHOLD HAS RISEN TO GREATER THAN 2.5 V AND IS CONSISTENTLY ABOVE 2.5 V.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD A LEAD WARNING FOR LOW IMPEDANCE. THE LEAD ALSO HAD UNDERSENSING AND WHEN TESTED THE LEAD HAD OVERSENSING AND NOISE WHEN THE PATIENT RAISED THEIR ARM. ADDITIONALLY, AN INSULATION BREACH WAS DETECTED. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. THE PATIENT WILL BE MONITORED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77748 CAPSURE Z NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5054-58

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention