HENRY SCHEIN ENDO-FLEX K-FILES 25 MM
Report
- Report Number
- 3007791573-2016-00001
- Date Received
- February 8, 2016
- Date of Event
- January 4, 2016
- Report Date
- February 5, 2016
- Manufacturer
- MANI HANOI CO., LTD.
- Product Code
- EKS
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- DENTIST
Narratives
AS WE COULD HAVE NOT RECEIVED THE ACTUAL PRODUCT OR THE INFORMATION REGARDING ITS LOT NUMBER, WE CHOSE A LOT NUMBER WHICH POSSIBLY HAD BEEN USED. WE EXECUTED BENDING TEST & TWISTING TEST. ITEM: ENDOFLEX #10K25. LOT NUMBER ITEM LOT# R1A02500-00 =MATERIAL LOT#R14B068004. AS A RESULT OF THAT, WE CONCLUDED THAT NO ABNORMALITY HAS BEEN IDENTIFIED. THE ABOVE REPORT HAS BEEN INFORMED TO HENRY SCHEIN ON MARCH 18, 2016.
THE EVENT WAS REPORTED BY THE IMPORTER ON 11TH OF JANUARY 2016 IN RESPONSE TO A CUSTOMER COMPLAIN. POST-MARKETING SURVEILLANCE MEETING WAS HELD AND IDENTIFIED AS NOT ADVERSE EVENT BASED ON THE PRODUCT RELATED TO THE EVENT WAS NOT LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY. TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED, AND CANNOT CONDUCT EVALUATION. ALSO LOT NUMBER IS NOT AVAILABLE. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. DEVICE NOT RETURNED TO MANUFACTURER.
AN IMPORTER REPORTED THAT WHILE A DENTIST WAS PERFORMING A ROOT CANAL PROCEDURE WITH K-FILE, IT BROKE AND GOT STUCK INSIDE THE PATIENT'S CANAL. THE DENTIST TOOK AN X-RAY AND WAS UNABLE TO RETRIEVE THE FILE; THEREFORE, THE PATIENT WAS REFERRED TO AN ORAL SURGEON TO HAVE IT REMOVED. THE BROKEN FILE WAS SUCCESSFULLY REMOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 75409 | HENRY SCHEIN ENDO-FLEX K-FILES 25 MM | FILE, PULP CANAL, ENDODONTIC | EKS | MANI HANOI CO., LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Other |