FDA Adverse Event Injury Summary report: N

MESO BIOMATRIX ACELLULAR PERITONEUM MATRIX

MDR report key: 5414566 · Received February 6, 2016

Report

Report Number
2530154-2016-00001
Event Type
Injury
Date Received
February 6, 2016
Date of Event
December 22, 2015
Report Date
February 5, 2016
Manufacturer
DSM BIOMEDICAL
Product Code
OXH
PMA / PMN Number
K094061
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED. THEREFORE, THE INVESTIGATION IS LIMITED TO REVIEW OF THE LOT HISTORY RECORDS. THE LOT HISTORY RECORDS INDICATE THAT THE DEVICES IN THE LOT MET ALL PRE-DETERMINED ACCEPTANCE CRITERIA. SEROMA AND INFECTION ARE KNOWN COMPLICATIONS THAT HAVE BEEN REPORTED IN THE LITERATURE TO OCCUR IN UP TO 20% OF IMPLANT-BASED BREAST RECONSTRUCTIONS. OUTCOME INFORMATION HAS BEEN REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT. IF ADDITIONAL INFORMATION IS PROVIDED, IT WILL BE REPORTED IN A FOLLOW-UP REPORT. DEVICE NOT RETURNED.

Description of Event or Problem · 1

A FEMALE PATIENT HAD BILATERAL, NIPPLE-SPARING MASTECTOMY FOLLOWED BY IMMEDIATE BREAST RECONSTRUCTION WITH BILATERAL MESO BIOMATRIX AND TISSUE EXPANDER IMPLANTATION. AT APPROXIMATELY 3 WEEKS POST-OPERATIVE, SHE REQUIRED ASPIRATION OF BILATERAL SEROMA AND BEGAN A 5 DAY COURSE OF ORAL ANTIBIOTICS. TWO DAYS LATER, MILD REDNESS WAS OBSERVED ON THE LOWER POLE OF THE LEFT BREAST. AT APPROXIMATELY 4 WEEKS POST-OPERATIVE, THE SEROMA RECURRED IN THE LEFT BREAST REQUIRING ASPIRATION. TWO DAYS LATER, A 'SLIGHT LEAK' WAS OBSERVED ALONG WITH CONTINUED REDNESS OF THE LEFT BREAST. INTRAVENOUS ANTIBIOTICS WERE STARTED AND SHE WAS SUBSEQUENTLY TAKEN TO THE OPERATING THEATER FOR SURGICAL EXPLORATION, WASHOUT AND VANCOMYCIN IRRIGATION OF THE LEFT BREAST. SHE WAS DISCHARGED 3 DAYS LATER AND BEGAN A 7 DAY COURSE OF ORAL ANTIBIOTICS. DURING THE FOLLOWING TWO WEEKS, THE LEFT BREAST SEEMED TO BE HEALING FINE. HOWEVER, REDNESS AND SWELLING OF THE LEFT BREAST RECURRED APPROXIMATELY ONE MONTH AFTER THE SURGICAL EXPLORATION. THE TISSUE EXPANDER AND MESO BIOMATRIX WERE SUBSEQUENTLY EXPLANTED FROM THE LEFT BREAST. AS OF THE DATE OF THIS REPORT, THE OUTCOME IS PENDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72678 MESO BIOMATRIX ACELLULAR PERITONEUM MATRIX MESH, SURGICAL, COLLAGEN, PLASTIC AND RECONSTRUCTIVE SURGERY OXH DSM BIOMEDICAL 30055-06 C5184

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention