FDA Adverse Event Malfunction Summary report: N

TRIAGE CARDIO PROFILER/TRIAGE CARDIAC PANEL

MDR report key: 5414010 · Received February 5, 2016

Report

Report Number
2027969-2016-00078
Event Type
Malfunction
Date Received
February 5, 2016
Date of Event
December 26, 2015
Report Date
January 15, 2016
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
NBC
PMA / PMN Number
K030286
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT CODE (CONTINUED): DDR, JHX AND MMI. INVESTIGATION/CONCLUSION: IT IS INDICATED THAT THE PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, A REVIEW OF THE IN-HOUSE TESTING HISTORY FOR LOT W60030RB WAS PERFORMED AND NO DISCREPANCIES WERE OBSERVED. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED AND THE LOT MET RELEASE SPECIFICATIONS. SINCE NO SAMPLE WAS RETURNED, SAMPLE SPECIFIC INTERFERENCE IS UNABLE TO BE RULED OUT AS A POTENTIAL CAUSE FOR DISCREPANT RESULTS. BASED ON THE INFORMATION AVAILABLE, THE PRODUCT PERFORMED AS EXPECTED. THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A VARIANCE BETWEEN THE TRIAGE TROPONIN (TNI) RESULT AND THE LABORATORY TNI RESULT. ON (B)(6) 2015, A (B)(6) FEMALE PRESENTED TO THE EMERGENCY ROOM WITH CHEST PAIN. THE TRIAGE TNI RESULT WAS 0.27 AT 12:14 AND THE LABORATORY TNI WAS <0.03 AT 13:00. BOTH RESULTS WERE FROM THE SAME SAMPLE DRAW. THE CUSTOMER HAD NO CONCERNS ABOUT SAMPLE QUALITY AS IT WAS CHECKED FOR HEMOLYSIS, CLOTS, LIPEMIA, ETC. ELECTROCARDIOGRAM (EKG) WAS PERFORMED BUT THE RESULT WAS NOT AVAILABLE PER CUSTOMER. ADDITIONAL LABORATORY TESTING AS FOLLOWS: (B)(6). THERE WERE NO INVASIVE PROCEDURES PERFORMED AND THE PATIENT WAS NOT HOSPITALIZED. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITH THE FOLLOWING DIAGNOSIS: CHEST WALL MUSCLE SPASM, HYPERTENSION, TOBACCO USE AND URINARY TRACT INFECTION (UTI).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71597 TRIAGE CARDIO PROFILER/TRIAGE CARDIAC PANEL CARDIAC MARKER TEST NBC ALERE SAN DIEGO, INC. 97022HS W60030RB

Patients

Seq Age Sex Outcome Treatment
1 49 YR MEDROXYPROGESTERONE (DEPO-PROVERA) 150MG/ML 3 MOS