FDA Adverse Event Malfunction Summary report: N

BOM 7MM EXTENDED LENGTH ENDOSCOPE

MDR report key: 5402249 · Received February 1, 2016

Report

Report Number
2242352-2016-00107
Event Type
Malfunction
Date Received
February 1, 2016
Report Date
January 6, 2016
Manufacturer
MAQUET CV
Product Code
GCJ
PMA / PMN Number
K014250
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS IS A REUSABLE OEM DEVICE; THEREFORE, A LOT HISTORY REVIEW WAS NOT APPLICABLE. THE CERTIFICATE OF CONFORMANCE (C OF C) WAS REVIEWED. THE VENDOR CERTIFIES THAT THIS DEVICE LOT CONFORMS TO ALL APPLICABLE PRODUCT SPECIFICATIONS. THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. SIGNS OF CLINICAL USE WERE OBSERVED. THERE WAS NO BLOOD OBSERVED ON THE DEVICE. THERE WERE NO DEFECTS OBSERVED IN THE VISUAL INSPECTION. AN IMAGE QUALITY INSPECTION WAS PERFORMED WITH AN ENDOSCOPIC VIDEO IMAGING SYSTEM. A CLEAR IMAGE WAS OBTAINED AND THE SIZE OF THE IMAGE WAS DETERMINED TO BE NORMAL. BASED ON THE RESULTS OF THE EVALUATION, THE REPORTED COMPLAINT WAS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

FEB. 01, 2016 01:51 PM (GMT-5:00) ADDED BY (B)(6): THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT THE BIO-MED DEPARTMENT RECEIVED A BOM 7MM EXTENDED LENGTH ENDOSCOPE WITH A NOTE STATING THE PICTURE ON THE MONITOR WAS TOO SMALL AND THAT SOMETHING MAY BE WRONG WITH THE SCOPE.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT THE BIO-MED DEPARTMENT RECEIVED A BOM 7MM EXTENDED LENGTH ENDOSCOPE WITH A NOTE STATING THE PICTURE ON THE MONITOR WAS TOO SMALL AND THAT SOMETHING MAY BE WRONG WITH THE SCOPE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61718 BOM 7MM EXTENDED LENGTH ENDOSCOPE LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ MAQUET CV 523265

Patients

Seq Age Sex Outcome Treatment
1