FDA Adverse Event
Malfunction
Summary report: N
SOLUTION WARMER
MDR report key: 5401402
·
Received February 1, 2016
Report
- Report Number
- 5401402
- Event Type
- Malfunction
- Date Received
- February 1, 2016
- Date of Event
- January 25, 2016
- Report Date
- January 27, 2016
- Manufacturer
- O.R. SOLUTIONS INC
- Product Code
- LHC
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
WILE CLEANING POST OPERATIVELY, NOTICED A HOLE ON THE DRAPE OF THE SOLUTION WARMER USED ON THE CASE WHEN SHE WAS DRAINING THE SOLUTION FROM IT. THE DRAPE APPEARED TO HAVE MELTED FROM THE HEAT. CLINICAL ENGINEERING PROVIDED THE FOLLOWING UPDATE: THIS EVENT IS RELATED TO AN OPERATING ROOM SOLUTIONS WARMER MODEL ORS2094RE. UNABLE TO DUPLICATE PROBLEM, RAN UNIT FOR 8 HOURS AT 120 DEGREES AND UNIT OPERATED TO MANUFACTURER'S SPECIFICATIONS. THE UNIT TESTED FINE DURING THE EVALUATION; HOWEVER, THIS IS THE SECOND TIME IN 5 MONTHS THAT THIS HAS HAPPENED WITH THIS PARTICULAR WARMER. RECOMMEND SENDING THE UNIT INTO OPERATING ROOM SOLUTIONS FOR A MORE COMPREHENSIVE TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60783 | SOLUTION WARMER | SOLUTION WARMER | LHC | O.R. SOLUTIONS INC | ORS2094RE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |