FDA Adverse Event Malfunction Summary report: N

SOLUTION WARMER

MDR report key: 5401402 · Received February 1, 2016

Report

Report Number
5401402
Event Type
Malfunction
Date Received
February 1, 2016
Date of Event
January 25, 2016
Report Date
January 27, 2016
Manufacturer
O.R. SOLUTIONS INC
Product Code
LHC
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

WILE CLEANING POST OPERATIVELY, NOTICED A HOLE ON THE DRAPE OF THE SOLUTION WARMER USED ON THE CASE WHEN SHE WAS DRAINING THE SOLUTION FROM IT. THE DRAPE APPEARED TO HAVE MELTED FROM THE HEAT. CLINICAL ENGINEERING PROVIDED THE FOLLOWING UPDATE: THIS EVENT IS RELATED TO AN OPERATING ROOM SOLUTIONS WARMER MODEL ORS2094RE. UNABLE TO DUPLICATE PROBLEM, RAN UNIT FOR 8 HOURS AT 120 DEGREES AND UNIT OPERATED TO MANUFACTURER'S SPECIFICATIONS. THE UNIT TESTED FINE DURING THE EVALUATION; HOWEVER, THIS IS THE SECOND TIME IN 5 MONTHS THAT THIS HAS HAPPENED WITH THIS PARTICULAR WARMER. RECOMMEND SENDING THE UNIT INTO OPERATING ROOM SOLUTIONS FOR A MORE COMPREHENSIVE TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60783 SOLUTION WARMER SOLUTION WARMER LHC O.R. SOLUTIONS INC ORS2094RE