ATTUNE FB TIBIAL IMPACTOR
Report
- Report Number
- 1818910-2016-12356
- Event Type
- Malfunction
- Date Received
- February 1, 2016
- Date of Event
- January 28, 2016
- Report Date
- January 29, 2016
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- HWA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED
CONCLUSION AND JUSTIFICATION STATUS: THE COMPLAINT STATES INTRAOPERATIVELY THE IMPACTORS BROKE. THE INVESTIGATION CONFIRMED THAT THE IMPACTORS HAD BROKEN AS REPORTED. EXPERT OPINION INDICATES THAT THE FAILURES ARE ASSOCIATED WITH ENVIRONMENTAL STRESS CRACKING (ESC). THE ATTUNE IMPACTORS HAVE BEEN ANNEALED TO REDUCE RESIDUAL STRESSES. LAB TESTS INDICATE SIGNIFICANT REDUCTION IN RESIDUAL STRESSES OF ANNEALED SAMPLES IN COMPARISON TO UN-ANNEALED SAMPLES HOWEVER A COMPLETE REDUCTION IN STRESSES IS NOT ACHIEVABLE. A FIELD SAFETY NOTICE WAS ISSUED IN (B)(6) 2014 STATING TO REDUCE THE POSSIBILITY OF LEAVING FRAGMENTS IN PATIENTS, THE COMPANY SUGGESTS ADHERING TO THE INSTRUCTIONS FOR USE (IFU), WHICH INCLUDE INSPECTING THE IMPACTOR¿S TO ENSURE THAT NO INSTRUMENTS OR PIECES OF INSTRUMENTS ARE LEFT IN THE SURGICAL SITE PRIOR TO CLOSURE. IT SHOULD BE NOTED THAT CAPA-003835 HAS BEEN INITIATED AND CAN BE REFERENCED FOR FURTHER DETAILS. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION; IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THE INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).
INTRAOPERATIVELY THE IMPACTOR BROKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59590 | ATTUNE FB TIBIAL IMPACTOR | KNEE INSTRUMENT/TRIAL | HWA | DEPUY ORTHOPAEDICS, INC. | AU3784954 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |