FDA Adverse Event Malfunction Summary report: N

ATTUNE FB TIBIAL IMPACTOR

MDR report key: 5401301 · Received February 1, 2016

Report

Report Number
1818910-2016-12356
Event Type
Malfunction
Date Received
February 1, 2016
Date of Event
January 28, 2016
Report Date
January 29, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
HWA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED

Additional Manufacturer Narrative · 1

CONCLUSION AND JUSTIFICATION STATUS: THE COMPLAINT STATES INTRAOPERATIVELY THE IMPACTORS BROKE. THE INVESTIGATION CONFIRMED THAT THE IMPACTORS HAD BROKEN AS REPORTED. EXPERT OPINION INDICATES THAT THE FAILURES ARE ASSOCIATED WITH ENVIRONMENTAL STRESS CRACKING (ESC). THE ATTUNE IMPACTORS HAVE BEEN ANNEALED TO REDUCE RESIDUAL STRESSES. LAB TESTS INDICATE SIGNIFICANT REDUCTION IN RESIDUAL STRESSES OF ANNEALED SAMPLES IN COMPARISON TO UN-ANNEALED SAMPLES HOWEVER A COMPLETE REDUCTION IN STRESSES IS NOT ACHIEVABLE. A FIELD SAFETY NOTICE WAS ISSUED IN (B)(6) 2014 STATING TO REDUCE THE POSSIBILITY OF LEAVING FRAGMENTS IN PATIENTS, THE COMPANY SUGGESTS ADHERING TO THE INSTRUCTIONS FOR USE (IFU), WHICH INCLUDE INSPECTING THE IMPACTOR¿S TO ENSURE THAT NO INSTRUMENTS OR PIECES OF INSTRUMENTS ARE LEFT IN THE SURGICAL SITE PRIOR TO CLOSURE. IT SHOULD BE NOTED THAT CAPA-003835 HAS BEEN INITIATED AND CAN BE REFERENCED FOR FURTHER DETAILS. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION; IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THE INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Description of Event or Problem · 1

INTRAOPERATIVELY THE IMPACTOR BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59590 ATTUNE FB TIBIAL IMPACTOR KNEE INSTRUMENT/TRIAL HWA DEPUY ORTHOPAEDICS, INC. AU3784954

Patients

Seq Age Sex Outcome Treatment
1