30CM PERCUTANEOUS ANTENNA X1
Report
- Report Number
- 3006451981-2016-00102
- Event Type
- Malfunction
- Date Received
- January 29, 2016
- Report Date
- September 18, 2015
- Manufacturer
- COVIDIEN MEDICAL PRODUCTS (SHANGHAI) MAN
- Product Code
- NEY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
COVIDIEN REFERENCE #: (B)(4). DATE OF INITIAL REPORT: (B)(6) 2016. THIS COMPLAINT IS PART OF A RETROSPECTIVE REVIEW AS PART OF A REMEDIATION RELATED TO (B)(4). COVIDIEN¿S INVESTIGATION IDENTIFIED TWO POSSIBLE ROOT CAUSES FOR THE DEVICE FAILURE. THE HUB DIVIDER SHIFTING DISTALLY WITHIN THE HUB OR THE INFLOW TUBE EXPANDING OR RUPTURING. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED. A SPACER WAS ADDED TO PREVENT THE HUB FROM SHIFTING. ADDITIONALLY, THE INFLOW TUBE MATERIAL WAS REVISED. THE NEW INFLOW TUBE IS MADE FROM POLYIMIDE, WHICH RESULTS IN INFLOW TUBES WITH MORE STRENGTH REDUCING THE LIKELIHOOD OF EXPANSION OR RUPTURE.
THE CUSTOMER REPORTED THAT AFTER 7 MINUTES, THE ANTENNA ALARMED AND STOPPED WORKING. THERE WAS NO SALINE FLOW AND ANTENNA GAVE HIGH TEMP ALERT. THE ANTENNA WAS REPLACED WITH A NEW ANTENNA AN THE CASE WAS SUCCESSFULLY COMPLETED. NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58592 | 30CM PERCUTANEOUS ANTENNA X1 | PERCUTANEOUS ANTENNA | NEY | COVIDIEN MEDICAL PRODUCTS (SHANGHAI) MAN | CA30L1 | S4LG010X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |