FDA Adverse Event Malfunction Summary report: N

30CM PERCUTANEOUS ANTENNA X1

MDR report key: 5399911 · Received January 29, 2016

Report

Report Number
3006451981-2016-00102
Event Type
Malfunction
Date Received
January 29, 2016
Report Date
September 18, 2015
Manufacturer
COVIDIEN MEDICAL PRODUCTS (SHANGHAI) MAN
Product Code
NEY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COVIDIEN REFERENCE #: (B)(4). DATE OF INITIAL REPORT: (B)(6) 2016. THIS COMPLAINT IS PART OF A RETROSPECTIVE REVIEW AS PART OF A REMEDIATION RELATED TO (B)(4). COVIDIEN¿S INVESTIGATION IDENTIFIED TWO POSSIBLE ROOT CAUSES FOR THE DEVICE FAILURE. THE HUB DIVIDER SHIFTING DISTALLY WITHIN THE HUB OR THE INFLOW TUBE EXPANDING OR RUPTURING. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED. A SPACER WAS ADDED TO PREVENT THE HUB FROM SHIFTING. ADDITIONALLY, THE INFLOW TUBE MATERIAL WAS REVISED. THE NEW INFLOW TUBE IS MADE FROM POLYIMIDE, WHICH RESULTS IN INFLOW TUBES WITH MORE STRENGTH REDUCING THE LIKELIHOOD OF EXPANSION OR RUPTURE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AFTER 7 MINUTES, THE ANTENNA ALARMED AND STOPPED WORKING. THERE WAS NO SALINE FLOW AND ANTENNA GAVE HIGH TEMP ALERT. THE ANTENNA WAS REPLACED WITH A NEW ANTENNA AN THE CASE WAS SUCCESSFULLY COMPLETED. NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58592 30CM PERCUTANEOUS ANTENNA X1 PERCUTANEOUS ANTENNA NEY COVIDIEN MEDICAL PRODUCTS (SHANGHAI) MAN CA30L1 S4LG010X