DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2016-15960
- Event Type
- Malfunction
- Date Received
- January 29, 2016
- Date of Event
- January 6, 2016
- Report Date
- January 6, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000187
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) DESCRIBE EVENT OR PROBLEM - ADDITIONAL. INITIAL REPORTER - CORRECTION, CORRECTION/ADDITIONAL INFORMATION/DEVICE EVALUATION.
(B)(4).
SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4) /LOT NUMBER 5201199), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON 02/20/2016. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT A PERMANENT OUT OF RANGE SIGNAL THAT OCCURRED ON (B)(6) 2015. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57445 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608 | 5202897 | 00386270000187 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |