Report
- Report Number
- 1818910-2016-12141
- Event Type
- Malfunction
- Date Received
- January 28, 2016
- Date of Event
- January 15, 2016
- Report Date
- January 18, 2016
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- HWR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
CONCLUSION AND JUSTIFICATION STATUS: THE COMPLAINT STATES ORTHOKIT RECEIVED A REPORT OF ITEM 920010029 LOT PC34371001 BROKE IN SURGERY. PLASTIC LOCKING SLEEVE USED TO FIX CUP ONTO IMPACTOR DID NOT WORK THE INVESTIGATION COULD NOT CONFIRM THE FAILURE MODE AS THE FULL DEVICE WAS NOT RETURNED. PREVIOUSLY A CORRECTIVE ACTION WAS IDENTIFIED FOR THIS FAILURE MODE WHEREBY A DESIGN CHANGE WAS IMPLEMENTED AND ROUTED ON ECO-193427 (DWG-106984) IN 2007 IN ORDER TO STRENGTHEN THE LOCKING CATCH. THE DESIGN VERIFICATION WAS COMPLETED AND ROUTED ON DVA-101743-FDE AND CONCLUDED THAT THERE ARE NO OUTSTANDING ACTIONS IDENTIFIED. THE COMPLAINT SHALL BE CLOSED WITH AN UNDETERMINED CONCLUSION IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
INSTRUMENT BROKE IN SURGERY, PLASTIC LOCKING SLEEVE USED TO FIX CUP ONTO IMPACTOR DID NOT WORK. THIS PART HAS BEEN THROWN AWAY BY THE HOSPITAL. NO DELAY OR HARM TO PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55930 | ANGLED ACET INSERTR | HIP INSTRUMENT/TRIAL | HWR | DEPUY ORTHOPAEDICS, INC. | PC34371001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |