FDA Adverse Event Injury Summary report: N

21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER

MDR report key: 5392246 · Received January 26, 2016

Report

Report Number
1024879-2016-00003
Event Type
Injury
Date Received
January 26, 2016
Date of Event
January 8, 2016
Report Date
January 12, 2016
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
PMA / PMN Number
D038055
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MEDICAL DEVICE LOT #: UNKNOWN. TWO POTENTIAL LOT NUMBERS WERE PROVIDED FOR THIS INCIDENT; 5235748 & 5230715. MEDICAL DEVICE EXPIRATION DATE: 5232748: 8/31/2020; 5230715: 8/31/2020. DEVICE MANUFACTURE DATE: 5232748: 8/20/2015; 5230715: 8/18/2015. RESULTS: A SAMPLE IS NOT AVAILABLE FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORDS REVEALED NO IRREGULARITIES FOR THE POTENTIAL LOT NUMBERS 5235748 & 5230715. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. HOWEVER, ALTHOUGH IT IS UNABLE TO BE CONFIRMED, OUR QUALITY ENGINEER NOTES THAT A POTENTIAL CAUSE FOR THIS INCIDENT MAY BE A VALVE FAILURE AT THE GEL DISPENSE OPERATION. A SAMPLE IS NOT AVAILABLE FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING A 21 G X 1.25 IN BD ECLIPSE BLOOD COLLECTION NEEDLE WITH LUER ADAPTER TO DRAW LABS, A PHLEBOTOMIST ACTIVATED THE SAFETY MECHANISM, HEARD AN AUDIBLE CLICK BUT STATED THAT THE NEEDLE WAS NOT FULLY COVERED BY THE SAFETY SHIELD. THIS RESULTED IN A CONTAMINATED NEEDLE STICK INJURY TO THE PHLEBOTOMIST. THE SOURCE PATIENT IS (B)(6) AND BOTH THE PATIENT AND THE PHLEBOTOMIST RECEIVE ROUTINE POST EXPOSURE LAB WORK. THE TESTS RESULTS WERE (B)(6) AND THE PHLEBOTOMIST WAS CURRENT ON HER (B)(6) IMMUNIZATION THUS NO ADDITIONAL MEDICAL TREATMENT OR PROPHYLAXIS WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
49425 21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER SAMPLE NEEDLE FMI BECTON, DICKINSON & CO., (BD) UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention