DIALOG A + HE/BIC
Report
- Report Number
- 3002879653-2016-00001
- Event Type
- Malfunction
- Date Received
- January 25, 2016
- Date of Event
- January 14, 2016
- Report Date
- January 15, 2016
- Manufacturer
- B. BRAUN AVITUM AG - MELSUNGEN
- Product Code
- FKJ
- PMA / PMN Number
- K083460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EXEMPTION NUMBER E2015012. B. BRAUN MEDICAL INC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN AVITUM AG (MANUFACTURER) THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN AVITUM INTERNAL REPORT (B)(4) (EVENT 1) ADDITIONAL INFORMATION HAS BEEN REQUESTED FROMN THE FACILITY AND THE INVESTIGATION IS ON-GOING AT THIS TIME. A FOLLOW UP REPORT WILL BE PROVIDED WHEN THE EVALUATION RESULTS ARE AVAILABLE.
EXEMPTION NUMBER E2015012. B. BRAUN MEDICAL INC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN AVITUM AG (MANUFACTURER) THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN AVITUM INTERNAL REPORT (B)(4) (EVENT 1) CUSTOMER REPORTED THAT A MACHINE PULLED TO MUCH FLUID DURING THERAPY ON TWO PATIENT TREATMENTS ON (B)(6) 2016. THE FIRST PATIENT SHOWED AN AMOUNT OF FLUID REMOVED GREATER THEN 1.3 LITER, WHILE ON THE SECOND PATIENT, THE AMOUNT WAS 1.5 LITER. THE CUSTOMER BIOMED RAN TWO MOCK THERAPIES. ON THE FIRST, HE SAW 1.2 L OVER FILTRATION. ON THE SECOND HE SAW 1.0 AND 1.5 L OVERFILTRATION. IN BOTH MOCK THERAPIES MACHINE WAS SET TO 1.5L (GOAL). THE MACHINE WAS REMOVED FROM SERVICE AND WAS LATER SERVICED BY A B. BRAUN AVITUM TECHNICIAN. THE TECHNICIAN REPLACED THE CONDUCTIVITY SENSORS AND RAN A MOCK THERAPY. MOCK THERAPY WAS SET FOR 2HRS WITH A UF GOAL OF 1 LITER. THE THERAPY REMOVED 997 ML, WELL WITHIN TOLERANCE. NO DEVIATIONS WERE NOTED AND THE MACHINE WAS RELEASED FOR OPERATION. THE SAMPLE AND ALL AVAILABLE INFORMATION WERE FORWARDED TO THE MANUFACTURER FOR FURTHER EVALUATION. THEIR INVESTIGATION REPORT STATES THAT THEY RECEIVED BIC AND END CONDUCTIVITY SENSORS AND CONFIRMED THE COMPLAINED ERROR. THERE WERE SOME CRACKS AT THE CONNECTION POINT TO THE DF-BLOCK. ACCORDING TO THEIR INVESTIGATION REPORT, THEY BECAME AWARE THAT A LIMITED NUMBER OF FLANGES ASSEMBLED INTO BICARBONATE AND END CONDUCTIVITY CELLS SHOWED HAIRLINE CRACKS. THE CONDUCTIVITY SENSORS ARE UTILIZED IN AN AREA OF THE DIALYSIS FLUID SYSTEM WHICH IS UNDER NEGATIVE PRESSURE. THE NEGATIVE PRESSURE ALLOWS AIR TO BE SUCKED INTO THE DIALYSIS FLUID. THE AIR COLLECTS IN AN AIR SEPARATOR WHICH FORMS PART OF THE DIALYSIS FLUID SYSTEM. WHEN THE AIR SEPARATOR IS FILLED, THE AIR SEPARATOR VALVE (VLA) OPENS AND THE AIR IS PURGED AND REPLACED BY DIALYSIS FLUID FROM OUTSIDE THE BALANCE CIRCUIT. THE LEAKAGE HAS DIFFERENT CONSEQUENCES DEPENDANT UPON WHETHER OR NOT THE DIALOG+ HEMODIALYSIS MACHINE IS EQUIPPED WITH A DF-FILTER OR NOT. FOR MACHINES WITH THE DF FILTER, THE AIR PASSES THE FILTER AND IS COLLECTED IN THE AIR SEPARATOR. THE LEAKAGE HAS DIFFERENT CONSEQUENCES DEPENDENT ON WHETHER A DIALOG+ MACHINE IS EQUIPPED WITH A DIALYSIS FLUID (DF) FILTER OR NOT. IN DIALOG MACHINES WITH DF-FILTER (OPTION DF-FILTER, HDF ONLINE MACHINES) THE AIR PASSES THIS FILTER ON ITS WAY TO THE AIR SEPARATOR. SINCE AIR BLOCKS THE FILTER IT IS FLUSHED OUT DURING CYCLIC FILTER RINSING, COLLECTED IN THE AIR SEPARATOR AND FLUSHED OUT AS DESCRIBED. CONTINUOUS AIR ENTRY DUE TO THE LEAKAGE OF THE CONDUCTIVITY CELLS CAN LEAD TO CYCLICAL PRESSURE FLUCTUATIONS DUE TO EXCESS AIR INSIDE THE DF-FILTER, THUS BLOCKING IT. THIS MIGHT LEAD TO AN EXCESS UF FOR THE PATIENT ("EXCESSIVE REMOVAL"). THE MAXIMUM POSSIBLE AMOUNT OF THE EXCESS UF (THEORETICAL WORST CASE) CAUSED BY THIS FAILURE TYPE WAS DETERMINED IN A LABORATORY SETTING ARTIFICIALLY PRODUCING THE MAXIMUM LEAKAGE AT WHICH THE DIALYSIS MACHINE CAN BE PUT IN THERAPY. AT A DIALYSIS FLOW OF 800 ML/MIN, THE UF DEVIATION WAS 636 ML/H WITH THE ALARM TRIGGERED AFTER 55 MIN OF THERAPY. IN ADDITION, 77 CONDUCTIVITY CELLS RETURNED FROM THE MARKET AND SHOWING THIS FAILURE TYPE WERE INVESTIGATED. 64 OF THESE CELLS SHOWED AN UF DEVIATION BELOW 200 ML/H AND 13 ABOVE 200 ML/H AT A DIALYSIS FLUID FLOW OF 800 ML/MIN. THE MAXIMUM UF DEVIATION MEASURED IN THIS CASE WAS 488 ML/H. IN DIALOG MACHINES WITHOUT DF-FILTER THE AIR IS ALSO TRANSPORTED TO THE AIR SEPARATOR, WHERE IT ACCUMULATES AND IS RELEASED AS DESCRIBED ABOVE. SINCE THERE IS NO DF-FILTER, IN THESE DEVICES NO ACCUMULATING AIR MIGHT CAUSE AN EXCESS UF. THE AIR ACCUMULATED IN THE AIR SEPARATOR EXPANDS WHEN IT IS SEPARATED AND SUBSTITUTED WITH DIALYSIS FLUID FROM OUTSIDE THE BALANCE CIRCUIT. SO THE FLUID VOLUME IS BIGGER THAN THE VOLUME OF THE COMPRESSED AIR AND THE UF VOLUME IS DECREASED. THEREFORE, IN DEVICES WITHOUT DF-FILTER AIR ENTRY MIGHT LEAD TO A TOO LOW ULTRAFILTRATION ("INSUFFICIENT REMOVAL"). THIS MECHANISM OCCURS ALSO IN DEVICES WITH DF-FILTER BUT IS SUPERIMPOSED BY THE DESCRIBED CYCLICAL PRESSURE FLUCTUATIONS INSIDE THE DF-FILTER. THE AMOUNT OF INSUFFICIENT UF CAUSED BY THIS FAILURE TYPE IN DIALYSIS MACHINES WITHOUT DF-FILTER WAS DETERMINED IN THE ABOVE MENTIONED THEORETICAL WORST CASE SETTING. THE MAXIMUM UF DEVIATION IN THIS CASE WAS 250 ML/H AT A DIALYSIS FLUID FLOW OF 800 ML/MIN. THE RESPECTIVE ALARM WAS TRIGGERED FOR THE FIRST TIME AFTER 13 MIN. TO FURTHER INVESTIGATE THE UF DEVIATION 24 CONDUCTIVITY CELLS RETURNED FROM THE MARKET WERE TESTED. THREE OF THESE CONDUCTIVITY CELLS SHOWED A DEVIATION BETWEEN 100 TO 200 ML/H WITH A MAXIMUM AT 148 ML/H, THE OTHER CONDUCTIVITY CELLS SHOWED LOWER OR NO UF DEVIATION. THE TESTS SHOWED THAT THE UF DEVIATIONS DECREASED WITH DECREASING DIALYSIS FLUID FLOW. THEY ALSO IMPLY THAT THE CRACKS AT THE CONDUCTIVITY CELLS DEVELOP AND INCREASE WITH TIME DURING CONTINUOUS USE OF THE DEVICE. WITH AN INCREASING LEAKAGE, THE AMOUNT OF AIR SUCKED INTO THE DIALYSIS FLUID SYSTEM INCREASES AND THE FREQUENCY OF THE OPENING OF THE VLA VALVE. THIS MEANS THAT THE ABOVE DESCRIBED ALARM STARTS TO OCCUR FIRST TOWARDS THE END OF A THERAPY AND OCCURS EARLIER AND EARLIER WITH THE GROWTH OF THE CRACK. IF THE LEAKAGE MIGHT APPROACH THE MAXIMUM, THE ALARM IS TRIGGERED EARLY IN THERAPY AND MORE FREQUENTLY. IF THE MAXIMUM IS REACHED, FOLLOW-UP ALARMS WILL STOP THE PREPARATION OR THERAPY PHASE. ROOT CAUSE: THE PRINCIPLE CAUSE OF FAILURE OF THE PRODUCT IS DUE TO VARIATION IN MOLDING PARAMETERS OF THE FLANGES UTILIZED IN THE CONSTRUCTION OF THE FOLLOWING CONDUCTIVITY SENSORS.DUE TO THE VARIATION IN MOLDING PARAMETERS, THERE IS POTENTIAL FOR THERMAL AND MECHANICAL STRESSES DURING NORMAL OPERATION TO INDUCE CRACKS IN THE CONDUCTIVITY SENSORS WHICH CAUSES THE INGRESS OF AIR INTO THE DIALYSIS FLUID. CORRECTION: B. BRAUN AVITUM AG (MANUFACTURER) AND B. BRAUN MEDICAL INC (IMPORTER) HAVE INITIATED AN URGENT MEDICAL DEVICE CORRECTION (RECALL) ON APRIL 1, 2016. ACTION TAKEN TO PREVENT RECURRENCE OF PROBLEM: B. BRAUN AVITUM AG (MANUFACTURER) HAS INITIATED CAPA2-2016-002. B. BRAUN AVITUM US (IMPORTER) HAS INITIATED CAPA2016-004-PFG. IF ADDITIONAL PERTINENT BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTED MACHINE PULLED TO MUCH FLUID DURING THERAPY ON TWO PATIENT TREATMENTS ON (B)(6) 2016.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46096 | DIALOG A + HE/BIC | HEMODIALYSIS SYSTEM | FKJ | B. BRAUN AVITUM AG - MELSUNGEN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |