15CM PERCUTANEOUS ANTENNA X1
Report
- Report Number
- 3006451981-2016-00072
- Event Type
- Malfunction
- Date Received
- January 22, 2016
- Date of Event
- October 27, 2015
- Report Date
- October 27, 2015
- Manufacturer
- COVIDIEN LLC (SHANGHAI)
- Product Code
- NEY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS COMPLAINT IS PART OF A RETROSPECTIVE REVIEW AS PART OF A REMEDIATION RELATED TO CAPA (B)(4) ALTHOUGH NO SAMPLE WAS RETURNED IN THIS CASE, IN OTHER SIMILAR INCIDENTS, COVIDIEN¿S INVESTIGATION IDENTIFIED TWO POSSIBLE ROOT CAUSES FOR THE DEVICE FAILURE. THE HUB DIVIDER SHIFTING DISTALLY WITHIN THE HUB OR THE INFLOW TUBE EXPANDING OR RUPTURING. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED. A SPACER WAS ADDED TO PREVENT THE HUB FROM SHIFTING. ADDITIONALLY, THE INFLOW TUBE MATERIAL WAS REVISED. THE NEW INFLOW TUBE IS MADE FROM POLYIMIDE, WHICH RESULTS IN INFLOW TUBES WITH MORE STRENGTH REDUCING THE LIKELIHOOD OF EXPANSION OR RUPTURE.
THE CUSTOMER REPORTED THE EMPRINT CA15L1 ANTENNA WAS USED TO ABLATE A COLORECTAL LIVER METS IN AN OPEN SURGICAL FIELD. THE 1ST ABLATION OF 2.5 MINS AT 100 WATTS WORKED PERFECTLY. THE ANTENNA WAS REMOVED AND RE-POSITIONED INTO A 2ND METS (THE GENERATOR AND PUMP WERE LEFT ON DURING THE RE-POSITIONING). THE WATTS WERE RESET TO 100 AND TIME TO 2 MINS, THE ABLATION COMMENCED AND AFTER 3 SECS A TEMP WARNING ALARM CAME ON AND THE PROCEDURE STOPPED. WE RESET AND TRIED AGAIN 4 MORE TIMES WITH THE SAME OUTCOME. THE PUMP WAS STOPPED AND RESET THE RESERVOIR TWICE AND NOTICED A POOR FLOW AND STILL GOT TEMPERATURE ALARM. THE SALINE BAG WAS SWAPPED FOR A NEW ONE AND RE-SET, AGAIN WITH THE SAME ALARM AFTER 2-3 SECONDS. A NEW ANTENNA WAS OPENED AND WE RESTARTED THE PROCEDURE, THIS TIME IT WORKED FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 44674 | 15CM PERCUTANEOUS ANTENNA X1 | ABLATION ANTENNA | NEY | COVIDIEN LLC (SHANGHAI) | S5GG009X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |