640G INSULIN PUMP
Report
- Report Number
- 3004209178-2016-82660
- Event Type
- Malfunction
- Date Received
- January 20, 2016
- Date of Event
- December 27, 2015
- Report Date
- June 1, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- Removal / Correction Number
- Z-0958-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PATIENT
- Health Professional
- N
Narratives
THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0958-2020 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
THE INSULIN PUMP PASSED FUNCTIONAL TESTING INCLUDING THE SELF-TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST. THE TEST GUARDIAN 2 LINK WITH THE GLUCOSE SENSOR SIMULATOR AND THE TEST BLOOD GLUCOSE METER COMMUNICATED PROPERLY TO THE INSULIN PUMP. ALL OF THE BUTTONS FUNCTIONED PROPERLY. NO DAMAGE IN THE KEYPAD ASSEMBLY NOTED. NO STUCK BUTTON ALARM DURING TESTING. NO UNLOCKED PRINTED CIRCUIT BOARD 1 KEYPAD CONNECTOR DURING OUR VISUAL INSPECTION. THE INSULIN PUMP PASSED THE LEAK TEST. THE INSULIN PUMP WAS RECEIVED WITH SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED RETAINER AND MINOR SCRATCHED LCD WINDOW.
THE CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE INSULIN PUMP ALARM STUCK BUTTON. CUSTOMER'S BLOOD GLUCOSE AT TIME OF INCIDENT WAS UNKNOWN MG/DL. IT WAS EXPLAINED TO THE CUSTOMER THE ALARM AND POSSIBLE CAUSES. THE CUSTOMER WAS ASKED TO CHECK DEVICE OPTIONS AND FOUNDER THERE WERE NO DEVICES ENTERED AT ALL. CUSTOMER STATED THAT THIS PROGRAMMED PREVIOUSLY. THE CUSTOMER STATED THEY ARE UNSURE IF THEY PRESSED A BUTTON DOWN FOR 3 MINUTES OR LONGER. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN. THE CUSTOMER WAS ADVISED THAT SINCE STUCK BUTTON LEADS TO PUMP REPLACEMENT WE DID NOT ATTEMPT TO RESOLVE RF ISSUES AND NO ALARMS IN HISTORY RELATING TO RF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38243 | 640G INSULIN PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1711K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |