HAKIM PROGRAMMABLE VALVE W/PRECHAMBER
Report
- Report Number
- 1226348-2016-10061
- Event Type
- Injury
- Date Received
- January 20, 2016
- Date of Event
- January 24, 2015
- Manufacturer
- CODMAN & SHURTLEFF, INC. / MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- PK974739
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
DEVICE EVALUATION: UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 120MMH2O. THE VALVE WAS VISUALLY INSPECTED, NO DEFECTS WERE NOTED. THE VALVE WAS IRRIGATED WITH PURIFIED WATER. NO OCCLUSION WAS NOTED. THE VALVE WAS DRIED. THE VALVE WAS LEAK TESTED, NO LEAKS NOTED. THE VALVE WAS REFLUX TESTED, THE VALVE PASSED THE TEST. THE VALVE WAS TESTED FOR PROGRAMMING. THE VALVE PASSED THE TEST. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3110 WITH LOT CPMBPR, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN (B)(6) 2013. NO ROOT CAUSE COULD BE DETERMINED AS THE VALVE FUNCTIONS. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.
(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.
CHECK FAILURE. THE VP VALVE WAS REMOVED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38868 | HAKIM PROGRAMMABLE VALVE W/PRECHAMBER | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. / MEDOS S.A. | CPMBPR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |