FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE W/PRECHAMBER

MDR report key: 5379846 · Received January 20, 2016

Report

Report Number
1226348-2016-10061
Event Type
Injury
Date Received
January 20, 2016
Date of Event
January 24, 2015
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
PK974739
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 120MMH2O. THE VALVE WAS VISUALLY INSPECTED, NO DEFECTS WERE NOTED. THE VALVE WAS IRRIGATED WITH PURIFIED WATER. NO OCCLUSION WAS NOTED. THE VALVE WAS DRIED. THE VALVE WAS LEAK TESTED, NO LEAKS NOTED. THE VALVE WAS REFLUX TESTED, THE VALVE PASSED THE TEST. THE VALVE WAS TESTED FOR PROGRAMMING. THE VALVE PASSED THE TEST. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3110 WITH LOT CPMBPR, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN (B)(6) 2013. NO ROOT CAUSE COULD BE DETERMINED AS THE VALVE FUNCTIONS. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.

Description of Event or Problem · 1

CHECK FAILURE. THE VP VALVE WAS REMOVED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
38868 HAKIM PROGRAMMABLE VALVE W/PRECHAMBER SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC. / MEDOS S.A. CPMBPR

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention