FDA Adverse Event Injury Summary report: N

PURELY YOURS ULTRA

MDR report key: 5375638 · Received January 18, 2016

Report

Report Number
3009974348-2016-00189
Event Type
Injury
Date Received
January 18, 2016
Date of Event
December 16, 2016
Report Date
December 21, 2016
Manufacturer
AMEDA, INC.
Product Code
HGX
PMA / PMN Number
K973501
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS EVALUATED FOR EVIDENCE OF ALLEGATION. THE RETURNED AMEDA PURELY YOURS BREAST PUMP MET AMEDA SPECIFICATIONS FOR BOTH SUCTION AND SPEED, AND PASSED VISUAL INSPECTION STANDARDS. NO EVIDENCE OF MALFUNCTION WAS OBSERVED. RETURNED PUMP PASSED ALL FUNCTIONAL TESTING.

Additional Manufacturer Narrative · 1

CUSTOMER WAS CONTACTED FOUR TIMES VIA EMAIL AND PHONE, REQUESTING RETURN OF THE BREAST PUMP TO AMEDA, INC. FOR INVESTIGATION. CUSTOMER RESPONDED, STATING SHE HAD A DEATH IN HER IMMEDIATE FAMILY AND WILL RETURN THE PUMP SOON. AS OF THIS DATE, PRODUCT HAS NOT BEEN RETURNED, THEREFORE NO VISUAL INSPECTION OR TESTING WAS PERFORMED ON THE PRODUCT. IF PRODUCT IS RETURNED TO AMEDA, INC., A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

CUSTOMER CONTACTED AMEDA, INC. ON (B)(6) 2015 BY EMAIL, REPORTING THE PURELY YOURS ULTRA BREAST PUMP BEGAN HAVING AN ISSUE WITH LOW SUCTION ON BOTH SIDES WHEN DUAL PUMPING, WITH THE LEFT SIDE HAVING LOWER SUCTION THAN THE RIGHT. CUSTOMER CONDUCTED SOME TROUBLESHOOTING ON HER OWN BY REPLACING PUMP PARTS WHICH DID NOT RESOLVE THE LOW SUCTION. CUSTOMER REPORTS LOW SUCTION LED TO DECREASED MILK OUTPUT, CLOGGED MILK DUCTS AND A DIAGNOSIS OF LEFT BREAST MASTITIS. CUSTOMER WAS PRESCRIBED A 10 DAY COURSE OF ORAL ANTIBIOTICS BY HER HEALTHCARE PROVIDER ON (B)(6) 2015. MASTITIS RESOLVED WITHIN 2 DAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33415 PURELY YOURS ULTRA ELECTRIC BREAST PUMP HGX AMEDA, INC. 24501879

Patients

Seq Age Sex Outcome Treatment
1 29 YR Other