FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 5375016 · Received January 18, 2016

Report

Report Number
3004753838-2016-15536
Event Type
Malfunction
Date Received
January 18, 2016
Date of Event
December 26, 2015
Report Date
December 26, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000125
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER WAS NOT RETURNED TO DEXCOM. THE RECEIVER (PART NUMBER STK-PR-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5202072), BEING USED WITH THE COMPLAINT TRANSMITTER, WAS RETURNED ON (B)(4) 2016. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE REPORTED EVENT OF AN INTERMITTENT OUT OF RANGE SIGNAL WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED AN INTERMITTENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33632 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9438-05 5203727 00386270000125

Patients

Seq Age Sex Outcome Treatment
1 9 YR