DIDECO COMPACTFLO EVOLUTION OXYGENATOR
Report
- Report Number
- 9680841-2015-00676
- Event Type
- Malfunction
- Date Received
- January 15, 2016
- Date of Event
- December 7, 2015
- Report Date
- December 18, 2015
- Manufacturer
- SORIN GROUP ITALIA
- Product Code
- DTZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE COMPLAINED DEVICE IS NOT DISTRIBUTED IN USA, BUT IT IS SIMILAR TO THE APEX HP M ADULT HOLLOW FIBER MEMBRANE OXYGENATOR DEVICE, WHICH IS DISTRIBUTED IN THE USA (510K #: K092895). ALTHOUGH THE COMPLAINED DEVICE IS AVAILABLE, THERE IS NO NEED FOR DEVICE INVESTIGATION SINCE THE ISSUE (BLOOD TO WATER LEAK) IS KNOWN AND A CAPA PROJECT (007/08) HAS BEEN COMPLETED. SORIN GROUP (B)(4) MANUFACTURES THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THE INCIDENT OCCURED IN (B)(6). (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT THAT BLOOD LEAKED INTO WATER LINES OF THE HEAT EXCHANGER OF THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THIS WAS DISCOVERED 30 MINUTES INTO THE PROCEDURE. THE CIRCUIT WAS CHANGED IN LESS THEN 3 MINUTES AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY. A REVIEW OF THE DHR COULD NOT IDENTIFY ANY DEVIATIONS OR NONCONFORMITIES RELEVANT TO THE ISSUE. THE COMPLAINED DEVICE WAS NOT RETURNED TO SORIN GROUP (B)(4) FOR INVESTIGATION. THIS IS A KNOWN ISSUE CAUSED BY CORROSION IN AREAS WITH METAL, WHICH ARE STRESSED BY MANUFACTURING PROCESSES. A CAPA PROJECT ((B)(4)) WAS COMPLETED IN APRIL 2015 TO REDUCE THE OCCURRENCE OF SIMILAR DEFECTS. THE HEAT EXCHANGER LOT OF THE COMPLAINED UNIT WAS ASSEMBLED BEFORE IMPLEMENTATION OF THE CORRECTIVE ACTIONS.
SORIN GROUP (B)(4) RECEIVED A REPORT THAT BLOOD LEAKED INTO WATER LINES OF THE HEAT EXCHANGER OF THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THIS WAS DISCOVERED 30 MINUTES INTO THE PROCEDURE. THE CIRCUIT WAS CHANGED IN LESS THEN 3 MINUTES AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 29457 | DIDECO COMPACTFLO EVOLUTION OXYGENATOR | OXYGENATOR, CARDIOPULMONARY BYPASS | DTZ | SORIN GROUP ITALIA | NA | 1503100154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |