FDA Adverse Event Malfunction Summary report: N

DIDECO COMPACTFLO EVOLUTION OXYGENATOR

MDR report key: 5370716 · Received January 15, 2016

Report

Report Number
9680841-2015-00676
Event Type
Malfunction
Date Received
January 15, 2016
Date of Event
December 7, 2015
Report Date
December 18, 2015
Manufacturer
SORIN GROUP ITALIA
Product Code
DTZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINED DEVICE IS NOT DISTRIBUTED IN USA, BUT IT IS SIMILAR TO THE APEX HP M ADULT HOLLOW FIBER MEMBRANE OXYGENATOR DEVICE, WHICH IS DISTRIBUTED IN THE USA (510K #: K092895). ALTHOUGH THE COMPLAINED DEVICE IS AVAILABLE, THERE IS NO NEED FOR DEVICE INVESTIGATION SINCE THE ISSUE (BLOOD TO WATER LEAK) IS KNOWN AND A CAPA PROJECT (007/08) HAS BEEN COMPLETED. SORIN GROUP (B)(4) MANUFACTURES THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THE INCIDENT OCCURED IN (B)(6). (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT THAT BLOOD LEAKED INTO WATER LINES OF THE HEAT EXCHANGER OF THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THIS WAS DISCOVERED 30 MINUTES INTO THE PROCEDURE. THE CIRCUIT WAS CHANGED IN LESS THEN 3 MINUTES AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY. A REVIEW OF THE DHR COULD NOT IDENTIFY ANY DEVIATIONS OR NONCONFORMITIES RELEVANT TO THE ISSUE. THE COMPLAINED DEVICE WAS NOT RETURNED TO SORIN GROUP (B)(4) FOR INVESTIGATION. THIS IS A KNOWN ISSUE CAUSED BY CORROSION IN AREAS WITH METAL, WHICH ARE STRESSED BY MANUFACTURING PROCESSES. A CAPA PROJECT ((B)(4)) WAS COMPLETED IN APRIL 2015 TO REDUCE THE OCCURRENCE OF SIMILAR DEFECTS. THE HEAT EXCHANGER LOT OF THE COMPLAINED UNIT WAS ASSEMBLED BEFORE IMPLEMENTATION OF THE CORRECTIVE ACTIONS.

Description of Event or Problem · 1

SORIN GROUP (B)(4) RECEIVED A REPORT THAT BLOOD LEAKED INTO WATER LINES OF THE HEAT EXCHANGER OF THE DIDECO COMPACTFLO EVOLUTION OXYGENATOR. THIS WAS DISCOVERED 30 MINUTES INTO THE PROCEDURE. THE CIRCUIT WAS CHANGED IN LESS THEN 3 MINUTES AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29457 DIDECO COMPACTFLO EVOLUTION OXYGENATOR OXYGENATOR, CARDIOPULMONARY BYPASS DTZ SORIN GROUP ITALIA NA 1503100154

Patients

Seq Age Sex Outcome Treatment
1 55 YR