FDA Adverse Event
Malfunction
Summary report: N
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
MDR report key: 5369042
·
Received January 14, 2016
Report
- Report Number
- 3004753838-2016-15422
- Event Type
- Malfunction
- Date Received
- January 14, 2016
- Date of Event
- December 21, 2015
- Report Date
- December 22, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000200
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED A PERMANENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Description of Event or Problem · 1
SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201199), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(4) 2016. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF AN PERMANENT OUT OF RANGE SIGNAL WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 24339 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608-BLU | 5201277 | 00386270000200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |