VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2016-00015
- Event Type
- Malfunction
- Date Received
- January 14, 2016
- Report Date
- December 15, 2015
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE RETURNED: DEVICE 1: MR290V; LOT 1505280304; DATE OF MANUFACTURE 28 MAY 2015; DEVICE 2: MR290V; LOT 1503090304; DATE OF MANUFACTURE 9 MARCH 2015; DEVICE 3: MR290V; LOT 1508100304; DATE OF MANUFACTURE 10 AUGUST 2015. METHOD: THREE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE THEY WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION OF THE RETURNED CHAMBERS REVEALED A CRACK IN THE CHAMBER DOME OF DEVICE 1. THE CRACK STARTS AT THE BASE AND STRETCHES UPWARDS. NO RESIDUE WAS FOUND AT THE CRACKS. NO REPORTABLE FAULTS WERE NOTED ON DEVICES 2 AND 3. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 150528. CONCLUSION: WE ARE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE DAMAGE OBSERVED ON THE RETURNED MR290V CHAMBER. HOWEVER, BASED ON THE INSPECTION CARRIED OUT IT IS LIKELY THAT THE RETURNED CHAMBERS WERE EXPOSED TO PRESSURES GREATER THAN 8KPA. THE INTEGRITY OF THE CHAMBER CAN BE AFFECTED WHEN PRESSURE EXCEED 8 KPA. EVERY MR290 CHAMBER IS PRESSURE TESTED FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. ANY CHAMBER WHICH FAILS EITHER OF THESE TESTS IS REJECTED. THE RETURNED CHAMBER WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: MAXIMUM OPERATING PRESSURE: 8KPA; USE OF THE MR290 ABOVE THE MAXIMUM OPERATING PRESSURE MAY LEAD TO CRACKING, WATER LEAKAGE AND, ON RARE OCCASIONS, COULD LEAD TO A LOSS OF VENTILATION PRESSURE; PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; DO NOT USE THE CHAMBER IF THE SEALS ARE NOT INTACT WHEN RECEIVED, OR IF IT HAS BEEN DROPPED.
A HEALTHCARE FACILITY IN (B)(6), REPORTED THAT THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS CRACKED AND WAS LEAKING WATER FROM THE BASE. NO PATIENT CONSEQUNCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 26783 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 1505280304 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |