FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 5355206 · Received January 8, 2016

Report

Report Number
3004753838-2016-40252
Event Type
Malfunction
Date Received
January 8, 2016
Date of Event
December 14, 2015
Report Date
December 15, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000019
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER WAS NOT RETURNED TO DEXCOM. THE RECEIVER (PART NUMBER STR-PR-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5201387), BEING USED WITH THE COMPLAINT TRANSMITTER, WAS RETURNED ON (B)(6) 2016. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT OF INACCURACIES. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2015. SENSOR WAS INSERTED ON (B)(6) 2015. ADDITIONALLY, IT WAS REPORTED THAT THE PATIENT DID NOT CALIBRATE AFTER EXPERIENCING INACCURACY. THE DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: IF THE DIFFERENCE BETWEEN YOUR SENSOR GLUCOSE READING AND BLOOD GLUCOSE VALUE IS GREATER THAN 20% OF THE BLOOD GLUCOSE VALUE FOR SENSOR GLUCOSE READINGS > 80 MG/DL OR GREATER THAN 20 POINTS FOR SENSOR GLUCOSE READINGS < 80 MG/DL, WASH YOUR HANDS AND TAKE ANOTHER BLOOD GLUCOSE MEASUREMENT. IF THE DIFFERENCE BETWEEN THIS SECOND BLOOD GLUCOSE MEASUREMENT AND THE SENSOR IS STILL GREATER THAN 20% FOR SENSOR GLUCOSE READINGS > 80 MG/DL OR GREATER THAN 20 POINTS FOR SENSOR GLUCOSE READINGS < 80 MG/DL, RECALIBRATE YOUR SENSOR USING THE SECOND BLOOD GLUCOSE VALUE. THE SENSOR GLUCOSE READING WILL CORRECT OVER THE NEXT 15 MINUTES. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12301 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9500-27 NI 00386270000019

Patients

Seq Age Sex Outcome Treatment
1 4 YR