FDA Adverse Event Malfunction Summary report: N

MPS2 MYOCARDIAL PROTECTION SYSTEM CONSOLE

MDR report key: 5352108 · Received January 7, 2016

Report

Report Number
1649914-2015-00118
Event Type
Malfunction
Date Received
January 7, 2016
Date of Event
December 8, 2015
Report Date
February 26, 2016
Manufacturer
QUEST MEDICAL, INC.
Product Code
DTR
UDI-DI
20634624521265
PMA / PMN Number
K041979
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION CONFIRMED THE ALLEGED TEMPERATURE ISSUE; HOWEVER, THE ALLEGED BACK-FLOW ALARMS COULD NOT BE DUPLICATED. THE CIRCULATION PUMP/MOTOR WAS STALLING WHICH IN TURN CAUSED THE TEMPERATURE ALARMS. THE CIRCULATION MOTOR WAS REPLACED AND THE UNIT RETURNED TO SERVICE. THE CAUSE OF THE BACK-FLOW ALARMS IS UNKNOWN.

Additional Manufacturer Narrative · 1

QUEST MEDICAL, INC. HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. QUEST MEDICAL, INC. DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE HOSPITAL PERFUSIONIST REPORTED AN ISSUE ENCOUNTERED WITH THE MPS2 CONSOLE DURING USE. THE REPORT STATED THAT THE CONSOLE HAD BEEN DISPLAYING ERROR CODES ("ARREST AGENT BACKFLOW" AND "ADDITIVE PRIME BACKFLOW") DURING RECENT CASES. THE REPORT ALSO STATED THAT THE CONSOLE DISPLAYED HEATER ALARMS AND SHUT OFF THE INTERNAL HEATER DURING THE LAST PROCEDURE. IN ALL INSTANCES THE PERFUSIONIST CONTINUED TO USE THE CONSOLE AND COMPLETED THE PROCEDURE WITH NO PATIENT COMPLICATIONS. A FIELD SERVICE ENGINEER (FSE) WAS ON SITE FOR PREVENTIVE MAINTENANCE WORK AND WAS ABLE TO EVALUATE THE CONSOLE AND CONFIRM THE ALARM CODES/ERRORS IN THE LOG DATA. THE CONSOLE WILL BE SHIPPED BACK TO THE MANUFACTURER FOR FURTHER EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10141 MPS2 MYOCARDIAL PROTECTION SYSTEM CONSOLE CPBP HEAT EXCHANGER DTR QUEST MEDICAL, INC. 5201260 20634624521265

Patients

Seq Age Sex Outcome Treatment
1 Disability