FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 5350268 · Received January 7, 2016

Report

Report Number
1220908-2015-03459
Event Type
Malfunction
Date Received
January 7, 2016
Report Date
December 21, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED DEFIB MONITOR FLEX CABLE. THE DEFIB MONITOR FLEX CABLE WAS REPLACED TO REMEDY THE MALFUNCTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT THE DEVICE FAILED SELF-TEST FOR DEFIB. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11645 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1