FDA Adverse Event Other Summary report: N

POLARIS ADJUSTABLE PRESSURE VALVE SHUNT

MDR report key: 5345165 · Received December 30, 2005

Report

Report Number
3001587388-2005-00384
Event Type
Other
Date Received
December 30, 2005
Date of Event
December 9, 2005
Report Date
December 30, 2005
Manufacturer
SOPHYSA
Product Code
GYK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
1

Narratives

Additional Manufacturer Narrative · 1

THE VALVE HAS BEEN RETURNED IN (B)(4) 2005. ANALYSIS HAS TO BE DONE. SEE SANNED PAGE.

Description of Event or Problem · 1

THIS POLARIS ADJUSTABLE PRESSURE VALVE WAS IMPLANTED TO A PT IN (B)(6). THE IMPLANTATION DEPTH WAS LESS THAN 10 MM FROM THE SKIN. IN EARLY (B)(6), THE NEUROSURGEON ATTEMPTED TO DECREASE THE SETTING PRESSURE OF THE VALVE TO 330 MMH2O ( OR 230 MMH2O) WITH AN ADJUSTMENT MAGNET AND A PRESSURE SELECTOR, IN VAIN. HE REPLACED THE VALVE WITH ANOTHER SPV-400 IN (B)(6). HE SUCCEEDED IN CHANGING THE PRESSURE OF THE EXPLANTED VALVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POLARIS ADJUSTABLE PRESSURE VALVE SHUNT SHUNT, CENTRAL NERVOUS SYSTEM & COMPONENTS GYK SOPHYSA POLARIS (R) VALVE P0322

Patients

Seq Age Sex Outcome Treatment
1 UNK Other