BD OBTURATOR FOR VENFLON¿/SAFELON¿
Report
- Report Number
- 1710034-2016-00001
- Event Type
- Injury
- Date Received
- January 4, 2016
- Date of Event
- December 22, 2015
- Report Date
- January 18, 2016
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS NOT MARKETED IN THE US. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
RESULTS: TWO UNUSED SAMPLES WERE RETURNED FOR EVALUATION. A VISUAL/MICROSCOPIC ANALYSIS REVEALED THAT BOTH UNITS WERE IN ONE PIECE AND THERE WAS NO MECHANICAL/PHYSICAL DAMAGE (SPLITS OR NICKS) TO THE THREADS ROD OR CONE OF THE OBTURATOR. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT NUMBER 5133626. CONCLUSION: AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS THERE WAS NO DEFINITE PHYSICAL OR MECHANICAL EVIDENCE THAT CONFIRMED AND SUPPORTED MANUFACTURING PROCESS RELATED ISSUES FOR THE DEFECT OBSERVED IN THIS INCIDENT. A POTENTIAL CAUSE FOR THIS FAILURE IS THE UNITS MAY HAVE BECOME JAMMED ON THE FEEDING SYSTEM. THE MODIFICATION TO THE MA66, OBTURATOR ASSEMBLY MACHINE VACUUM PICK-UP HEAD, WAS IMPLEMENTED ON 4/2015. THE TAIL CLIPPING ISSUE WITH THE PICK-UP HEAD WAS A KNOWN POTENTIAL PROBLEM IN THE PAST AND WAS ADDRESSED WITH A MODIFICATION TO THE PICK-UP HEAD BEING THE SOLUTION.
IT WAS REPORTED THAT WHILE REMOVING A BD OBTURATOR FOR VENFLON⯓AFELON⠆ROM AN IV CATHETER, THE DEVICE BROKE OFF AND REMAINED INSIDE THE PATIENT. THE PATIENT HAD SURGERY TO TRY AND RETRIEVE THE BROKEN OBTURATOR BUT THE DEVICE WAS NOT FOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264 | BD OBTURATOR FOR VENFLON¿/SAFELON¿ | OBTURATOR | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 5133626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |