FDA Adverse Event Injury Summary report: N

BD OBTURATOR FOR VENFLON¿/SAFELON¿

MDR report key: 5341699 · Received January 4, 2016

Report

Report Number
1710034-2016-00001
Event Type
Injury
Date Received
January 4, 2016
Date of Event
December 22, 2015
Report Date
January 18, 2016
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS NOT MARKETED IN THE US. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

RESULTS: TWO UNUSED SAMPLES WERE RETURNED FOR EVALUATION. A VISUAL/MICROSCOPIC ANALYSIS REVEALED THAT BOTH UNITS WERE IN ONE PIECE AND THERE WAS NO MECHANICAL/PHYSICAL DAMAGE (SPLITS OR NICKS) TO THE THREADS ROD OR CONE OF THE OBTURATOR. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT NUMBER 5133626. CONCLUSION: AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS THERE WAS NO DEFINITE PHYSICAL OR MECHANICAL EVIDENCE THAT CONFIRMED AND SUPPORTED MANUFACTURING PROCESS RELATED ISSUES FOR THE DEFECT OBSERVED IN THIS INCIDENT. A POTENTIAL CAUSE FOR THIS FAILURE IS THE UNITS MAY HAVE BECOME JAMMED ON THE FEEDING SYSTEM. THE MODIFICATION TO THE MA66, OBTURATOR ASSEMBLY MACHINE VACUUM PICK-UP HEAD, WAS IMPLEMENTED ON 4/2015. THE TAIL CLIPPING ISSUE WITH THE PICK-UP HEAD WAS A KNOWN POTENTIAL PROBLEM IN THE PAST AND WAS ADDRESSED WITH A MODIFICATION TO THE PICK-UP HEAD BEING THE SOLUTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE REMOVING A BD OBTURATOR FOR VENFLON⯓AFELON⠆ROM AN IV CATHETER, THE DEVICE BROKE OFF AND REMAINED INSIDE THE PATIENT. THE PATIENT HAD SURGERY TO TRY AND RETRIEVE THE BROKEN OBTURATOR BUT THE DEVICE WAS NOT FOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264 BD OBTURATOR FOR VENFLON¿/SAFELON¿ OBTURATOR FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5133626

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention