FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM

MDR report key: 5340796 · Received January 4, 2016

Report

Report Number
2028159-2016-00009
Event Type
Malfunction
Date Received
January 4, 2016
Report Date
January 25, 2016
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K120912
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).

Additional Manufacturer Narrative · 1

THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS NOT REPLICATED. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A REVIEW OF THE CUSTOMER¿S COMPLAINT HISTORY FOR THE LAST 24 MONTHS DID NOT SHOW ANY PREVIOUS COMPLAINTS OF THIS KIND AGAINST THE SYSTEM. THE SYSTEM WAS MANUFACTURED ON JULY 25, 2005. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS. THEREFORE, THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE ESTABLISHED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED EXPERIENCING ASPIRATION ISSUES BEFORE THE PROCEDURE BEGAN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
746 INFINITI VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER NA NA

Patients

Seq Age Sex Outcome Treatment
1 Other