FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CONNECTING CABLE

MDR report key: 533513 · Received March 12, 2004

Report

Report Number
1418479-2004-00006
Event Type
Malfunction
Date Received
March 12, 2004
Date of Event
February 10, 2004
Report Date
February 13, 2004
Manufacturer
BOWA ELECTRONICS GMBH
Product Code
IKD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A MONOPOLAR ELECTRO-SURGICAL CABLE WAS IN USE DURING A LAPAROSCOPIC CHOLECYSTECTOMY CASE. WHEN POWER WAS APPLIED, THE CABLE BURNED IN TWO NEAR THE END THAT CONNECTED TO THE GENERATOR. ANOTHER CABLE WAS SUBSTITUTED IMMEDIATELY TO COMPLETE THE CASE. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CONNECTING CABLE MONOPOLAR CONNECTING CABLE IKD BOWA ELECTRONICS GMBH 8106.033 UNK