LIGHT HEAD E 668 NC W/O PAD
Report
- Report Number
- 0001220865-2015-00010
- Event Type
- Malfunction
- Date Received
- December 28, 2015
- Date of Event
- December 2, 2015
- Report Date
- December 2, 2015
- Manufacturer
- BERCHTOLD CORP
- Product Code
- FSY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
IT WAS REPORTED THAT A LIGHTHEAD ALLEGEDLY FELL OFF OF A SPRING ARM. THIS OCCURRED AFTER A CASE AND DURING CLEAN UP. A STRYKER FIELD SERVICE TECHNICIAN (SFST) EVALUATED THE ISSUE AND FOUND THAT THE BRAKE SCREWS WERE LOOSE. THE SFST FIXED THE ISSUES AND ALL FUNCTIONS WERE NORMAL. NO ADVERSE EVENTS OR INJURIES WERE REPORTED. A SUPPLEMENTAL MDR WILL BE FILED IF DEEMED NECESSARY PENDING CONCLUSION OF THE STRYKER QUALITY ENGINEER'S INVESTIGATION.
IT WAS REPORTED THAT AN E668 LIGHT HEAD FELL FROM THE SPRING ARM. THIS OCCURRED AFTER A CASE DURING CLEAN UP. THE LIGHT HEAD DID NOT FALL TO THE GROUND, IT WAS CAUGHT BY INTERNAL WIRING. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES. WHEN THE BIOMED AT THE SITE EXAMINED THE STATE OF THE EQUIPMENT, HE DETERMINED THE RETAINING SCREW AND THE BRAKE SCREW WERE BOTH PARTIALLY BACKED OUT. THE ROOT CAUSE OF THIS FAILURE IS BELIEVED TO BE CUSTOMER MISUSE IN REMOVING THESE SCREWS.
IT WAS REPORTED THAT A LIGHTHEAD ALLEGEDLY FELL OFF OF A SPRING ARM. THIS OCCURRED AFTER A CASE AND DURING CLEAN UP. NO ADVERSE EVENTS OR INJURIES WERE REPORTED.
IT WAS REPORTED THAT A LIGHTHEAD ALLEGEDLY FELL OFF OF A SPRING ARM. THIS OCCURRED AFTER A CASE AND DURING CLEAN UP. NO ADVERSE EVENTS OR INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 854271 | LIGHT HEAD E 668 NC W/O PAD | CHROMOPHARE E668 | FSY | BERCHTOLD CORP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |