FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 5326615
·
Received December 24, 2015
Report
- Report Number
- 2032227-2015-78322
- Event Type
- Malfunction
- Date Received
- December 24, 2015
- Date of Event
- October 13, 2015
- Report Date
- December 10, 2015
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. MOTOR PASSED MOTOR TEST. HOWEVER, THE INSULIN PUMP PASSED IDLE CURRENT TEST, RUN CURRENT TEST, SELF-TEST, OFF NO POWER TEST, UNEXPECTED RESTART ERROR TEST, BASIC OCCLUSION TEST, DISPLACEMENT TEST. THE INSULIN PUMP WAS RECEIVED WITH MISSING SEGMENTS DUE TO CRACKED LCD ZEBRA CONNECTOR. THE INSULIN PUMP WAS HAD MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
THE INSULIN PUMP WAS RETURNED FOR FAILURE ANALYSIS. THE REASON FOR THE RETURN WAS UNKNOWN. THE CUSTOMER'S BLOOD GLUCOSE WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 851770 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LWWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |