FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 5326615 · Received December 24, 2015

Report

Report Number
2032227-2015-78322
Event Type
Malfunction
Date Received
December 24, 2015
Date of Event
October 13, 2015
Report Date
December 10, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARM DURING OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. MOTOR PASSED MOTOR TEST. HOWEVER, THE INSULIN PUMP PASSED IDLE CURRENT TEST, RUN CURRENT TEST, SELF-TEST, OFF NO POWER TEST, UNEXPECTED RESTART ERROR TEST, BASIC OCCLUSION TEST, DISPLACEMENT TEST. THE INSULIN PUMP WAS RECEIVED WITH MISSING SEGMENTS DUE TO CRACKED LCD ZEBRA CONNECTOR. THE INSULIN PUMP WAS HAD MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE INSULIN PUMP WAS RETURNED FOR FAILURE ANALYSIS. THE REASON FOR THE RETURN WAS UNKNOWN. THE CUSTOMER'S BLOOD GLUCOSE WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
851770 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LWWS

Patients

Seq Age Sex Outcome Treatment
1