OT VERIO2 METER
Report
- Report Number
- 2939301-2015-55696
- Event Type
- Malfunction
- Date Received
- December 24, 2015
- Report Date
- December 14, 2015
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW-UP # 2: THE LAY USER/PATIENT'S TEST STRIPS HAVE BEEN RETURNED AND FURTHER EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE EVALUATED AND FOUND TO BE MISCLASSIFIED BY THE METER. THE METER REPORTED ¿CONTROL SOLUTION¿ WHEN BLOOD HAS BEEN APPLIED TO A STRIP. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
FOLLOW-UP # 1.THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS FAILED TESTING. THE REPORTED ISSUE WAS CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ BLOOD SAMPLES AS A CONTROL SOLUTION RESULTS. THE REPORTER CLAIMED BLOOD GLUCOSE TESTS WERE PERFORMED AND RESULTS OF "166, 205, 84, 84, 171 93, 274, 221, 114, 99, 99 AND 300 MG/DL" WERE OBTAINED WHICH THE METER FLAGGED AS CONTROL SOLUTION TESTS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 851275 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3879655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |