FDA Adverse Event Other Summary report: N

BD LATITUDE DIABETES MANAGEMENT SYSTEM

MDR report key: 532429 · Received March 5, 2004

Report

Report Number
2243072-2004-00140
Event Type
Other
Date Received
March 5, 2004
Date of Event
February 7, 2004
Report Date
March 4, 2004
Manufacturer
BECTON DICKINSON
Product Code
CFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT METER GIVING INACCURATE READINGS, WHEN COOMPARED TO THEIR OTHER METER. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READINGS (125, 211, 142, 280) AS REF. POINT VS. BD READINGS 353, 389, 346, 548 YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LATITUDE DIABETES MANAGEMENT SYSTEM BD TEST STRIPS CFR BECTON DICKINSON NA 2057336

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other