FDA Adverse Event
Other
Summary report: N
BD LATITUDE DIABETES MANAGEMENT SYSTEM
MDR report key: 532429
·
Received March 5, 2004
Report
- Report Number
- 2243072-2004-00140
- Event Type
- Other
- Date Received
- March 5, 2004
- Date of Event
- February 7, 2004
- Report Date
- March 4, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- CFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT METER GIVING INACCURATE READINGS, WHEN COOMPARED TO THEIR OTHER METER. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READINGS (125, 211, 142, 280) AS REF. POINT VS. BD READINGS 353, 389, 346, 548 YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LATITUDE DIABETES MANAGEMENT SYSTEM | BD TEST STRIPS | CFR | BECTON DICKINSON | NA | 2057336 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |