FDA Adverse Event Malfunction Summary report: N

CALCANEAL COMPRESSION SCREW, RIGHT A3 NAIL M10 X 10MM

MDR report key: 5320112 · Received December 22, 2015

Report

Report Number
0008031020-2015-00625
Event Type
Malfunction
Date Received
December 22, 2015
Date of Event
June 2, 2014
Report Date
June 3, 2014
Manufacturer
STRYKER GMBH
Product Code
HSB
PMA / PMN Number
K130147
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING DATE WAS ADDED.

Additional Manufacturer Narrative · 1

WITH GUIDANCE FROM THE MDR POLICY BRANCH OF THE FDA, MDR REPORTED BY STRYKER (B)(4) AS A RESULT OF A RETROSPECTIVE LOOKBACK OF COMPLAINTS RESULTING FROM THE ACQUISITION OF ASSETS FROM SMALL BONE INNOVATION, INC. (B)(4). DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE IT WILL BE PROVIDED ON A SUPPLEMENTAL REPORT. THE DEVICE IS NOT AVAILABLE TO STRYKER.

Description of Event or Problem · 1

LEFT/RIGHT COMPRESSION SCREWS WERE LABELED (LASER ETCHED) INCORRECTLY. THE LEFT SCREWS FIT INTO RIGHT NAILS AND THE RIGHT SCREWS FIT INTO LEFT NAILS. COMPRESSION SCREWS FOR LOT 1331001 FOR PART NUMBERS 933-0033 AND 933-0034 WERE CONFIRMED TO BE ASSEMBLED INCORRECTLY. THIS LOT OF PARTS FOR BOTH PART NUMBERS WAS REWORKED, RESTORING PARTS TO THEIR INTENDED CONFIGURATION OF LEFT HAND AND RIGHT HAND.

Description of Event or Problem · 1

LEFT/RIGHT COMPRESSION SCREWS WERE LABELED (LASER ETCHED) INCORRECTLY. THE LEFT SCREWS FIT INTO RIGHT NAILS AND THE RIGHT SCREWS FIT INTO LEFT NAILS. COMPRESSION SCREWS FOR LOT 1331001 FOR PART NUMBERS 933-0033 AND 933-0034 WERE CONFIRMED TO BE ASSEMBLED INCORRECTLY. THIS LOT OF PARTS FOR BOTH PART NUMBERS WAS REWORKED, RESTORING PARTS TO THEIR INTENDED CONFIGURATION OF LEFT HAND AND RIGHT HAND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
845737 CALCANEAL COMPRESSION SCREW, RIGHT A3 NAIL M10 X 10MM ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES HSB STRYKER GMBH 1331001

Patients

Seq Age Sex Outcome Treatment
1 Other