FDA Adverse Event
Other
Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 5316323
·
Received September 30, 2004
Report
- Report Number
- 2243072-2004-00597
- Event Type
- Other
- Date Received
- September 30, 2004
- Date of Event
- August 30, 2004
- Report Date
- September 30, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT HIS METER IS NOT SHOWING ACCURATE READINGS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (174) AS REF. POINT VS. BD READINGS (57, 242, 223) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 3064079 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |