FDA Adverse Event Malfunction Summary report: N

SPECIAL EVD 10-110 W/O Y SITE

MDR report key: 5314336 · Received December 21, 2015

Report

Report Number
2648988-2015-00079
Event Type
Malfunction
Date Received
December 21, 2015
Date of Event
December 3, 2015
Report Date
December 3, 2015
Manufacturer
INTEGRA NEUROSCIENCES PR
Product Code
JXG
PMA / PMN Number
K062599
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCTS WERE NOT RETURNED FOR EVALUATION. DEVICE HISTORY RECORD (DHR) OF THE FINISH GOOD (FG) LOT # 1151963 WAS REVIEWED. THE FG LOT # 1151963 WAS RELEASED FOR DISTRIBUTION ON JUNE 17, 2015 IN COMPLIANCE WITH THE PRODUCT SPECIFICATIONS AND INTEGRA REQUIREMENTS. NO ANOMALY OR DISCREPANCIES WERE REPORTED DURING THE MANUFACTURE OF THE FG LOT. NO OTHER COMPLAINTS HAVE BEEN REPORTED FOR FG LOT # 1151963 RELATED TO HAIR FOUND INSIDE PACKAGED PRODUCT OR ANY OTHER CONDITION. (B)(4). GIVEN THAT THE COMPLAINT UNIT HAS NOT BEEN RECEIVED FOR EVALUATION, THE REPORTED CONDITION (HAIR INSIDE OF PACKAGE) COULD NOT BE CONFIRMED. IT WAS REPORTED IN THE COMPLAINT RECORD THAT THE HAIR WAS FOUND INSIDE THE STERILE PACKAGE. THE LOCATION OF THE HAIR SUGGESTS THAT IT MOST LIKELY ENTERED THE POUCH DURING THE PACKAGING OPERATION.

Description of Event or Problem · 1

A HAIR WAS FOUND IN THE STERILE PACKAGE. THERE WAS NO PATIENT CONTACT, INJURY, OR DELAY IN SURGERY. THERE WAS NO PATIENT PREPPED FOR SURGERY. NO DELAY IN SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
842341 SPECIAL EVD 10-110 W/O Y SITE N/A JXG INTEGRA NEUROSCIENCES PR 1151963

Patients

Seq Age Sex Outcome Treatment
1