FDA Adverse Event Malfunction Summary report: N

NAVLOCK TRACKER UNIVERSAL GREY

MDR report key: 5313793 · Received December 21, 2015

Report

Report Number
1723170-2015-01573
Event Type
Malfunction
Date Received
December 21, 2015
Date of Event
November 23, 2015
Report Date
June 2, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
OLO
PMA / PMN Number
K124004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. DEVICE LOT NUMBER NOT AVAILABLE. DEVICE MANUFACTURING DATE IS DEPENDENT ON LOT NUMBER, THEREFORE, UNAVAILABLE. RETURN REQUESTED. REPLACEMENT DEVICE SHIPPED TO SITE 12/11/2015. NO PARTS HAVE BEEN RECEIVED BY MANUFACTURER FOR ANALYSIS.

Additional Manufacturer Narrative · 1

DEVICE LOT NUMBER NOW PROVIDED.DEVICE MANUFACTURING DATE NOW PROVIDED.

Additional Manufacturer Narrative · 1

ALTHOUGH THE COMPLAINT ALLEGES A SINGLE TRACKER WAS DAMAGED, 3 TOTAL TRACKERS WERE RETURNED FOR ANALYSIS: LOT# 100521- MANUFACTURER DUE: 05/21/2010 AND 2 OF LOT # 150814 - MANUFACTURE DATE: 08/14/2015. MEDTRONIC INVESTIGATION OF RETURNED SUSPECT TRACKER FINDS THAT THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE RETURNED TRACKERS WERE FOUND TO BE IN GOOD CONDITION WITH NO APPARENT PHYSICAL DAMAGE. KNOWN GOOD INSTRUMENTS INSERTED AND ROTATED WITHOUT ISSUE. WITH MARKERS ATTACHED AND FULLY, THE TRACKERS RETURNED GOOD GEOMETRY AND DIVOT ERROR READINGS. NO PROBLEM FOUND.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED A SITE'S GRAY NAVLOCK TRACKER WAS DAMAGED. WHILE USING THE NON-MEDTRONIC POWEREASE SYSTEM, THE TRACKER AND THE TAP BECAME JAMMED TOGETHER. THE TAP APPEARED TO BE STUCK ON THE TRACKER. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
839954 NAVLOCK TRACKER UNIVERSAL GREY ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 150814

Patients

Seq Age Sex Outcome Treatment
1