FDA Adverse Event
Malfunction
Summary report: N
POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
MDR report key: 5313404
·
Received December 21, 2015
Report
- Report Number
- 1525712-2015-05859
- Event Type
- Malfunction
- Date Received
- December 21, 2015
- Report Date
- December 11, 2015
- Manufacturer
- INVACARE TAYLOR STREET
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A SUPPLEMENTAL RECORD WILL BE FILED.
Additional Manufacturer Narrative · 1
ADDITIONAL/UPDATED INFORMATION WAS ADDED TO REFLECT THE DEVICE BEING RETURNED TO THE MANUFACTURER FOR EVALUATION. THE RESULT OF THE EVALUATION WAS THAT THERE WAS AN INDUCTIVE FAILURE OF THE JOYSTICK CAUSING THE REPORTED COMPLAINT ISSUE.
Description of Event or Problem · 1
JOYSTICK INTERMITTENLY MOVES THE CHAIR WITHOUT COMMAND, ALSO THE WINDOWS (OR DRIVES) ON THE JOYSTICK WILL CHANGE JUST BY MOVING THE JOYSTICK KNOB AND THEN THE JOYSTICK WILL GO A JOYSTICK FAULT.
Description of Event or Problem · 1
JOYSTICK INTERMITTENTLY MOVES THE CHAIR WITHOUT COMMAND, ALSO THE WINDOWS (OR DRIVES) ON THE JOYSTICK WILL CHANGE JUST BY MOVING THE JOYSTICK KNOB AND THEN THE JOYSTICK WILL GO A JOYSTICK FAULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 841521 | POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI | WHEELCHAIR, POWERED | ITI | INVACARE TAYLOR STREET | 3GAR-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |