FDA Adverse Event Injury Summary report: N

EQUIMATRIX CANCELLOUS (8503-06-0025)

MDR report key: 5310333 · Received December 18, 2015

Report

Report Number
2410375-2015-00013
Event Type
Injury
Date Received
December 18, 2015
Report Date
December 15, 2015
Product Code
DZE
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ON (B)(6) 2015, AN INITIAL, SPONTANEOUS REPORT WAS RECEIVED BY LUITPOLD PHARMACEUTICALS, INC FROM A DENTIST CONCERNING A FEMALE PATIENT (AGE, MEDICAL HISTORY, CONCOMITANT MEDICATIONS NOT PROVIDED). ON UNSPECIFIED DATE, THE DENTIST IMPLANTED EQUIMATRIX CANCELLOUS FOR GRAFTING IMPLANT (DOSE, LOT NUMBER NOT PROVIDED). ON AN UNSPECIFIED DATE, THE PATIENT COMPLAINED OF GETTING FOOD STUCK AROUND THE TEMPORARY IMPLANT AND DID NOT MAKE FOR A GOOD FIT FOR THE TEMPORARY. UPON EXAM, THE DENTIST NOTED THE TISSUE WAS NOT TIGHT AGAIN'S THE TEMPORARY. THE DENTIST INFORMED THE PATIENT, THE FOOD OR CRACKER WAS SHOVED (PER THE DENTIST) IN THE GUM BETWEEN THE TEMPORARY AND THE FACIAL GUM. THE PROBE WENT 10 MM DEEP. THE TEMPORARY WAS REMOVED AND CLEANED THE EMPTY SPACE OF THE TEMPORARY. THE SPACE DID NOT RESOLVE AND IT THE NEXT VISIT IT HAD WORSENED. THE DENTIST REPORTED THAT HE TOOK THE FOTONA LIGHT AND STERILIZED THE BUCCAL THREADS WITH ED-YAG AND GRAFT MFDBA (MINERALIZED FREEZE DRIED BONE ALLOGRAFT). THE TISSUE LOOKS NORMAL AND ADHERING TO THE TEMPORARY. THE DENTIST STATED THAT HE REMOVED ALL OF THE EQUIMATRIX BY THE ER-YAG WATER EXPLOSIONS. THE DENTIST REPORTED THIS AS A GRAFT FAILURE (MEDDRA: DEVICE FAILURE). COMPANY ASSESSMENT: INCREASED IMPLANT AND GRAFT FAILURE RATES CAN BE EXPECTED IN PATIENTS EXHIBITING RISK FACTORS SUCH AS SYSTEMIC DISEASES, SMOKERS, CHRONIC PERIODONTITIS, AND PERIAPICAL PATHOLOGY. IN ADDITION, CERTAIN ANATOMIC FACTORS INCLUDING THIN TISSUE BIOTYPE, BUCCAL WALL DEFECTS, AND POOR BONE DENSITY ARE IMPORTANT DETERMINANTS OF RISK ASSESSMENT. THE FOLLOWING COMMENTS APPLY TO ALL EQUIMATRIX CASES REPORTED BY THE DENTIST. THE CAUSALITY ASSESSMENT OF ALL CASES IS HAMPERED DUE TO A LACK OF DETAILED INFORMATION: NO PATIENT IDENTIFIERS WERE FURNISHED NO PATIENT DEMOGRAPHIC INFORMATION, SOCIAL HISTORY, OR HABIT HISTORY WAS PROVIDED NO INFORMATION REGARDING PATIENTS' MEDICAL HISTORY, PAST DENTAL HISTORY, MEDICATIONS OR CURRENT MEDICAL STATUS THAT MIGHT INFLUENCE NORMAL WOUND HEALING WAS PROVIDED NO CLINICAL OR RADIOGRAPHIC DIAGNOSIS (EXCEPT AS NOTED IN INDIVIDUAL REPORTS) NO DESCRIPTION OF DETAILS OF TREATMENT RENDERED NO SPECIFICATION OF CONCOMITANT MEDICAL DEVICES USED DURING SURGERY (EXCEPT AS NOTED) SUCH AS OTHER GRAFT MATERIALS, IMPLANT TYPE, MEMBRANES, ETC. NO DETAILS OF THE SURGICAL PROCEDURES WERE PROVIDED AND THERE WAS NO INFORMATION ABOUT HOW GRAFT MATERIALS WERE PROTECTED DURING HEALING NO COMMENTS ON OCCLUSAL ANALYSIS OR OCCLUSAL STATUS AND PARAFUNCTIONAL HABITS WAS PROVIDED RADIOGRAPHS WERE NOT PROVIDED FOR THE MAJORITY OF CASES. WHEN PROVIDED, THEY WERE NOT LABELED OR DATED (EXCEPT AS NOTED) CBCT IMAGES PROVIDED WERE SCREEN SHOTS THAT WERE NOT LABELED AND IN SOME CASES NOT DATED. PRE- AND POST-SURGERY CBCT IMAGES WERE FROM DIFFERENT SECTIONS AND COULD NOT BE COMPARED. RADIOGRAPHIC AND CBCT IMAGES DID NOT COINCIDE WITH DESCRIPTION PROVIDED BY THE DENTIST IN THE ACCOMPANYING E-MAILS TOOTH NUMBER AND IMPLANT SITES WERE NOT SPECIFIED FOR CASES AND SITES WITH COMPLICATIONS WERE NOT ALWAYS IDENTIFIED. BASED ON THE COMPANY REPRESENTATIVE'S TELEPHONE CONVERSATION WITH THE DENTIST ON (B)(6) 2015, NO FLAPS WERE REFLECTED PRIOR TO INITIATING ANY SURGERY POTENTIALLY LIMITING A THOROUGH DIRECT CLINICAL EXAMINATION FOLLOWING TOOTH REMOVAL AS WELL AS POSSIBLY COMPROMISING A THOROUGH POSTEXTRACTION DEBRIDEMENT OF EACH TREATED SITE. ACCORDING TO THE PACKAGE INSERT, "IN ORDER TO ASSURE THE FORMATION OF NEW BONE, EQUIMATRIX SHOULD ONLY BE PLACED IN DIRECT CONTACT WITH WELL VASCULARIZED BONE. CORTICAL BONE SHOULD BE MECHANICALLY PERFORATED, IF NECESSARY." WITHOUT REFLECTING A FLAP, IT WOULD BE IMPOSSIBLE TO DETERMINE WHETHER THE EQUIMATRIX WAS BEING PLACED IN DIRECT CONTACT WITH WELL-VASCULARIZED BONE, ESPECIALLY IF SIGNIFICANT SOCKET WALL DEFECTS WERE PRESENT WITHOUT REFLECTING A FLAP, IT WOULD BE DIFFICULT, AT BEST, TO: DETERMINE WHETHER THE INNER SOCKET WALLS REQUIRE MECHANICAL PERFORATIONS AS WELL AS MAKING SUCH PERFORATIONS EXCEEDINGLY DIFFICULT TO CARRY OUT. IF CORTICAL PERFORATION IN ORDER TO INCREASE THE BLOOD SUPPLY TO THE GRAFTED AREAS WERE NEEDED AND POSSIBLE, SUCH PERFORATION WOULD NEED TO BE ACCOMPLISHED PRIOR TO IMPLANT PLACEMENT. THERE IS NO INDICATION THAT ANY CORTICAL PERFORATION WAS CARRIED OUT PRIOR TO IMPLANT PLACEMENT FOR ANY OF THE CASES DISCUSSED BY THE DENTIST. ACCORDING TO THE PACKAGE INSERT, "AFTER EXPOSURE OF THE BONY DEFECT WITH A FULL THICKNESS MUCOPERIOSTEAL FLAP, ALL GRANULATION TISSUE MUST BE CAREFULLY REMOVED." THEREFORE, ACCORDING TO THE EQUIMATRIX PACKAGE INSERT, AND SUPPORTED BY NUMEROUS CLINICAL PHOTOS FOUND IN THE EQUIMATRIX BROCHURE, A FULL-THICKNESS FLAP IS REQUIRED PRIOR TO EQUIMATRIX GRAFT PARTICLE PLACEMENT. ACCORDING TO THE DENTIST, ALL CASES GRAFTED WITH EQUIMATRIX WERE PERFORMED "FLAPLESS". ACCORDING TO THE PACKAGE INSERT AND INCLUDED IN THE EQUIMATRIX BROCHURE SENT TO CLINICIANS, "SITES GRAFTED WITH EQUIMATRIX SHOULD BE ALLOWED TO HEAL APPROXIMATELY 6 MONTHS PRIOR TO IMPLANT PLACEMENT." IN OTHER WORDS, IMMEDIATE IMPLANT PLACEMENT FOLLOWING TOOTH REMOVAL IS NOT AN APPROVED INDICATION WHEN USING EQUIMATRIX. OF THE THREE CASES DISCUSSED IN SOME DETAIL, ALL WERE IMMEDIATELY PLACED IMPLANTS GRAFTED WITH EQUIMATRIX. THERE IS NO INFORMATION ABOUT THE PRE-EXTRACTION DIAGNOSIS FOR EACH AREA TREATED. WHY THE TEETH REQUIRED REMOVAL IN THE FIRST PLACE COULD IMPACT ON THE FINAL POST-IMPLANT AND GRAFTING RESULTS BOTH IN THE SHORT AND LONG-TERM. IT APPEARS THAT ALL IMPLANTS IN THE REPORTED CASES RECEIVED IMMEDIATE TEMPORARY RESTORATIONS. IT IS NOT KNOWN IF PREMATURE LOADING OF THE IMPLANTS WAS A FACTOR IN THE GRAFT AND IMPLANT FAILURE. IN THIS CASE, THE DENTIST DESCRIBES POST-OPERATIVE COMPLICATIONS AFTER USING EQUIMATRIX AROUND AN IMMEDIATE IMPLANT WHERE HE PLACED AN IMMEDIATE TEMPORARY IMPLANT CROWN. HE SAID THAT AFTER REMOVING THE EQUIMATRIX, HE DISINFECTED THE IMPLANT THREADS WITH A LASER AND THEN RE-GRAFTED WITH ALLOGRAFT. HE CONTENDS THAT THE REMOVAL OF THE EQUIMATRIX RESOLVED THE PROBLEM. FROM THE AVAILABLE INFORMATION, THERE IS NO CAUSE AND EFFECT BETWEEN THE INITIAL POOR TISSUE RESPONSE AND THE EQUIMATRIX. IT IS MORE LIKELY THAT THE ADVERSE OUTCOME IS DUE TO IMPROPER CLINICIAN PERFORMANCE (INCLUDING REPEATED REMOVAL OF THE TEMPORARY CROWN DURING EARLY HEALING) AND NOT DUE TO THE EQUIMATRIX PRODUCT. MANUFACTURER PRODUCT ASSESSMENT: LUITPOLD PHARMACEUTICALS IN COLLABORATION WITH (B)(4), THE MANUFACTURER OF EQUIMATRIX HAS CONDUCTED AN INVESTIGATION INTO THE EQUIMATRIX CUSTOMER COMPLAINTS. THIS INVESTIGATION INCLUDED A REVIEW OF THE MANUFACTURING PROCESS FOR THE FOLLOWING PRODUCT BATCHES: (B)(4). THIS REVIEW INCLUDED AN IN-DEPTH REVIEW OF THE BATCH RECORDS AND A CRITICAL ANALYSIS INCLUDING STERILITY TEST, PROTEIN ASSAY AND ENDOTOXIN TEST, WHICH MAY BE RELATED TO THE EFFECTIVENESS OF THE PRODUCTS. THE BATCH RECORD REVIEW INDICATED THAT ALL MANUFACTURING PROCEDURES COMPLIED WITH THE MANUFACTURING SOP WITH NO DEVIATION. THE CRITICAL BIOLOGIC ANALYSIS PERFORMED ON THE FOLLOWING BATCHES: (B)(4) DEMONSTRATED THAT THE PRODUCT IS MAINTAINED STERILE, THE PROTEIN CONTENT IS WELL BELOW THE CRITICAL LIMIT, AND ENDOTOXIN WAS NOT DETECTED. BASED ON THE INVESTIGATION CONDUCTED, IT CAN BE CONCLUDED THAT NO ABNORMALITIES COULD BE IDENTIFIED IN THE MANUFACTURING PROCESS AND THIS PRODUCT HAS BEEN DEMONSTRATED TO CONFORM TO THE CURRENT APPROVED SPECIFICATION. A COMPLETE COMPANY CAUSALITY ASSESSMENT AND MANUFACTURER PRODUCT QUALITY ASSESSMENT HAVE BEEN ATTACHED FOR FURTHER REVIEW. UPON INTERNAL REVIEW, IT WAS DETERMINED TO SUBMIT THIS CASE AS NON-EXPEDITED REPORT. THIS CASE IS LINKED TO (B)(4) SAME REPORTER AND (B)(4) CASE REPORT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
836367 EQUIMATRIX CANCELLOUS (8503-06-0025) NATURAL BONE MINERAL MATRIX DZE UNK

Patients

Seq Age Sex Outcome Treatment
1