FDA Adverse Event Malfunction Summary report: N

BONE PIN- 4MM X 80MM- STERILE 2 PACK

MDR report key: 5309798 · Received December 18, 2015

Report

Report Number
3005985723-2015-00307
Event Type
Malfunction
Date Received
December 18, 2015
Date of Event
December 16, 2015
Report Date
December 16, 2015
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K112507
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

REPORTED EVENT: THE 3.2X80MM BONE PINS DID NOT ACHIEVE GOOD FIXATION WITH THE BONE AT FIRST ATTEMPT. A SINGLE BONE PIN WAS PULLED BACK OUT AND IT WAS NOTED THAT THE TIP WAS NOT THERE. WE OPEN ANOTHER PACK OF PINS AND CONTINUED WITH THE CASE. ONLY ONE OF THE BONE PINS HAD THE TIP BROKE OFF OUT OF THE FIRST PACK. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (143000-01 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED SHIPPED TO (B)(4) ON JULY 2, 2015 AND ACCEPTED INTO FINAL STOCK ON JULY 2, 2015 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 143080, LOT NUMBER W41375 SHOWS THREE COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THE RELATED PI NUMBERS ARE (B)(4). TRACKING OF COMPLAINTS RELATED TO THE 143080 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST #(B)(4). CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF THE BONE PIN. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, 206388 REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD. DEVICE WAS NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) . DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL..

Description of Event or Problem · 1

THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) . DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
836918 BONE PIN- 4MM X 80MM- STERILE 2 PACK STEREOTACTIC DEVICE, ACCESSORY OLO MAKO SURGICAL CORP. W41375-1

Patients

Seq Age Sex Outcome Treatment
1 Other