FDA Adverse Event Malfunction Summary report: N

SMALL BATTERY DRIVE 12V BATTERY

MDR report key: 5306766 · Received December 17, 2015

Report

Report Number
8030965-2015-12355
Event Type
Malfunction
Date Received
December 17, 2015
Date of Event
December 4, 2015
Report Date
December 7, 2015
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

CORRECTION: THE LOT NUMBER WAS INADVERTENTLY ENTERED AS SERIAL NUMBER ON THE INITIAL REPORT. THE LOT NUMBER WAS UPDATED AS D150819 AND THE SERIAL NUMBER AS NA. DEVICE EVALUATION: THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND OBSERVED THAT THE BATTERY COULD NOT CHARGE BECAUSE THE CONTACTS WERE NOT WELDED PROPERLY. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO FAULTY PRODUCTION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT BEFORE SURGERY, IT WAS OBSERVED THAT THE BATTERY DEVICE WAS NOT CHARGING. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
835074 SMALL BATTERY DRIVE 12V BATTERY INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF D150819

Patients

Seq Age Sex Outcome Treatment
1