FDA Adverse Event Other Summary report: N

BDE PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 5302862 · Received September 17, 2004

Report

Report Number
2243072-2004-00561
Event Type
Other
Date Received
September 17, 2004
Date of Event
August 17, 2004
Report Date
September 17, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.

Description of Event or Problem · 1

CONSUMER REPORTS HAVING LOW READINGS WITH HER PLINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (179) AS REF POINTS VS BD READINGS (32) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BDE PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064077

Patients

Seq Age Sex Outcome Treatment
1 UNK Other