FDA Adverse Event Malfunction Summary report: N

COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET

MDR report key: 5296959 · Received December 15, 2015

Report

Report Number
3002808486-2015-00166
Event Type
Malfunction
Date Received
December 15, 2015
Report Date
August 10, 2017
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
PMA / PMN Number
K090140
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF# (B)(4). CATALOG # UNKNOWN AS INFORMATION WAS NOT PROVIDED BUT REFERRED TO AS A COOK CELECT FILTER. AS CATALOG # IS UNKNOWN IT COULD BE EITHER K061815, K073374, K090140, K1112119, K121057 OR K121629. (B)(4). IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4) CORRECTED DATA BASED ON NEW INFORMATION RECEIVED: ADVERSE EVENT TO PRODUCT PROBLEM, SERIOUS INJURY TO MALFUNCTION. THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016032. WILLIAM COOK (B)(4) (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURER REFERENCE # (B)(4). IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING ¿VENA CAVA PERFORATION, DEVICE IS UNABLE TO BE RETRIEVED". COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2014". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2014 AT THE (B)(6)". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION WAS RECEIVED ON 01/06/2016 AS FOLLOWS: PLAINTIFF ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2014 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DVT WITH MASSIVE BILATERAL PE. A CT REPORT FROM (B)(6) 2015 DESCRIBES THE FILTER LEGS PENETRATING THE WALL OF THE IVC AND THE MEDIAL LEG CONTACTING BUT NOT PROTRUDING INTO THE ABDOMINAL AORTA. NO RETRIEVAL ATTEMPTS WERE NOTED. PLAINTIFF IS ALLEGING VENA CAVA PERFORATION, DEVICE IS UNABLE TO BE RETRIEVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
828333 COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening