DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-77838
- Event Type
- Malfunction
- Date Received
- December 14, 2015
- Date of Event
- November 19, 2015
- Report Date
- November 19, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
PATIENT CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED A PERMANENT OUT OF RANGE SIGNAL. THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. THE DATA WAS REVIEWED ON (B)(6) 2015 AND CONFIRMED THE REPORTED EVENT OF INTERMITTENT ANTENNA ICON. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201330), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(6) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 822450 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22495-PNK | 5201358 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |