DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-73892
- Event Type
- Malfunction
- Date Received
- December 14, 2015
- Date of Event
- November 18, 2015
- Report Date
- November 18, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT OF A FIRMWARE ERROR. A ROOT CAUSE COULD NOT BE DETERMINED. ADDITIONALLY, THE TRANSMITTER (PART NUMBER STT-GL-003/(B)(4), LOT NUMBER 5190526) BEING USED WITH THE RECEIVER WAS RETURNED FOR EVALUATION. THE TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE DEXCOM G4® PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM REV. 11, UNDER SECTION 13.8.2 RECEIVER ERROR CODE NOTES THE FOLLOWING: THIS SCREEN SHOWS AN ERROR CODE THAT MEANS THE RECEIVER MAY NOT BE WORKING PROPERLY. WRITE DOWN THE ERROR CODE AND CONTACT DEXCOM TECHNICAL SUPPORT. CONTINUE TO CHECK YOUR BLOOD GLUCOSE VALUE USING YOUR BLOOD GLUCOSE METER. NO ALERT SOUND OR VIBRATION WILL WARN YOU THAT YOU ARE NO LONGER GETTING SENSOR GLUCOSE READINGS.
PATIENT CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT THE RECEIVER DISPLAYED A FIRMWARE ERROR ON (B)(6) 2015. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 822272 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22495 | 5204949 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |