FDA Adverse Event
Other
Summary report: N
MEVITRON 74
MDR report key: 52856
·
Received November 25, 1996
Report
- Report Number
- 52856
- Event Type
- Other
- Date Received
- November 25, 1996
- Date of Event
- October 1, 1996
- Report Date
- November 21, 1996
- Manufacturer
- SIEMEN'S MEDICAL SYSTEMS
- Product Code
- IYE
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- OH, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PT CAME FOR THERAPUTIC RADIATION TREATMENT. DOSAGE RATES SET. STUDENT WAS ABLE TO SET THE MACHINE ON HI ELECTRON MODE WITHOUT WARNING THE OPERATOR THAT THE HIGH ELECTRON DOSAGE WAS ON, PT RECEIVED 2100 CGY, RATHER THAN 300 CGY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEVITRON 74 | LINEAR THERAPEUTIC RADIATION | IYE | SIEMEN'S MEDICAL SYSTEMS | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Other |