FDA Adverse Event Injury Summary report: N

PANOSOL 3D

MDR report key: 5284950 · Received December 7, 2015

Report

Report Number
1521608-2015-00005
Event Type
Injury
Date Received
December 7, 2015
Date of Event
November 5, 2015
Report Date
December 1, 2015
Manufacturer
NATIONAL BIOLOGICAL CORP.
Product Code
FTC
PMA / PMN Number
K872651
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR'S OFFICE REPORTED THAT THE PATIENT CONFIRMED THAT HE TREATED HIMSELF WITHOUT READING THE OPERATIONS MANUAL, AND DID NOT ADHERE TO THE TREATMENT GUIDELINES SET FORTH BY HIS DOCTOR. PER THE INSTRUCTION MANUAL AND THE GUIDELINES OF THE PHYSICIAN, THE MAXIMUM TREATMENT EXPOSURE SHOULD NOT HAVE EXCEEDED 2 MINUTES AND 13 SECONDS. THE DOCTOR'S OFFICE CONFIRMED THAT THE PATIENT TREATED FOR 10 MINUTES ON BOTH THE FRONT OF HIS BODY AND ANOTHER 10 MINUTES ON THE BACK OF HIS BODY. NBC REQUESTED ACCESS TO THE DEVICE TO CONDUCT AN EVALUATION; HOWEVER, THE REQUEST FOR ACCESS WAS DENIED. THE DEVICE HAD JUST BEEN DELIVERED TO THE PATIENT, SO NBC REVIEWED THE DHR AND TEST RECORDS FOR THE DEVICE AND WAS ABLE TO CONFIRM THAT THE DEVICE WAS FUNCTION AS INTENDED 76 DAYS BEFORE THE EVENT. THE DOCTOR'S OFFICE AND NBC STRONGLY BELIEVE THAT THE PATIENT INJURY WAS CLEARLY A RESULT OF MISUSE OF INSTRUCTIONS FOR USE AS IT PERTAINS TO TREATMENT PROTOCOL.

Description of Event or Problem · 1

NATIONAL BIOLOGICAL CORPORATION RECEIVED A CALL FROM A DOCTOR'S OFFICE REGARDING A PATIENT INJURY. THE PATIENT WAS USING PANOSOL 3D SYSTEM, MODEL; UVB-631 AND WAS BURNED. THE EXACT LOCATION AND DEGREE OF BURNS ARE NOT KNOWN AT THIS TIME, BUT THE DOCTOR'S OFFICE DID INFORM NBC THAT THE PATIENT WAS TREATED IN THE ICU. CURRENT STATUS OF THE PATIENT WAS NOT AVAILABLE AT THE TIME OF THIS REPORT, BUT NBC WAS ABLE TO CONFIRM THAT THE PATIENT WAS RELEASED HOME DURING A FOLLOW-UP CALL WITH THE DOCTOR'S OFFICE, ((B)(6) 2015).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
801872 PANOSOL 3D FTC NATIONAL BIOLOGICAL CORP. UVB-631 10005

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention