FDA Adverse Event
Other
Summary report: N
FLURO-ETHYL SKIN REFRIGERENTGEBAUER'S ETHYL CHLORIDE
MDR report key: 527984
·
Received May 11, 2004
Report
- Report Number
- 527984
- Event Type
- Other
- Date Received
- May 11, 2004
- Date of Event
- April 1, 2004
- Report Date
- April 1, 2004
- Manufacturer
- GEBAUER CO
- Product Code
- MLY
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
A MD (MEDICAL DOCTOR) APPLIED ETHYL CHLORIDE (LOCAL ANESTHESIA) SPRAY TO PT'S FOREHEAD. ONCE THE AREA DRIED, THE PHYSICIAN TURNED ON AN ELECTRIC CAUTERY DEVICE TO I&D (INCISION AND DRAINAGE) A BOIL-LIKE ABSCESS. THE PT'S SYNTHETIC HAIR CAUGHT ON FIRE. THE FIRE WAS DOUSED BY A BLANKET AND NORMAL SALINE WAS USED TO SATURATE THE PT'S HEAD. A FIRST DEGREE BURN WAS SUSTAINED ON THE TOP OF THE PT'S EAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLURO-ETHYL SKIN REFRIGERENTGEBAUER'S ETHYL CHLORIDE | TOPICAL ANESTHETIC SKIN REFRIGERANT | MLY | GEBAUER CO | * | * | |
| 2 | BATTERY OPERATED CAUTERIES | CAUTERY | HQP | ABCO DEALERS, INC | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR | Other | ON PATIENT: NO OTHER "DEVICE, FLAMMABLE TOPICAL| CAUTERY FOR I&, D OF ABSCESS OTHER DEVICES IN| ANESTHESIA SPRAY, ANTIBIOTICS. |